Optimum configuration of cannulated hip screws for the fixation of intracapsular hip fractures: a biomechanical study

被引:126
作者
Selvan, VT
Oakley, MJ
Rangan, A
Al-Lami, MK
机构
[1] Middlesbrough Gen Hosp, Dept Trauma & Orthopaed, Middlesbrough, Cleveland, England
[2] Univ Durham, Dept Biomech Engn, Durham DH1 3HP, England
来源
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED | 2004年 / 35卷 / 02期
关键词
D O I
10.1016/S0020-1383(03)00059-7
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
The radiographs of a sample of patients who had canulated hip screw fixation for intracapsular femoral neck fractures were reviewed in our region. There were six different types of configurations used in these fixations which are divided into two groups: (I) triangular configurations, consisting of two parallel screws with a third screw placed either superiorly, inferiorly, anteriorly or posteriorly; and (II) linear configurations with two or three screws in a vertical line. In our study, we tested the relative strength of each configuration in a laboratory setting using synthetic bone models. Statistical analysis, at 5% significance level, using two-way ANOVA and post-hoc test was carried out to test the differences of the results between the configurations. Our results clearly show that the triangular configurations had a higher peak load, higher ultimate toad, less displacement and more energy absorption before failure than other configurations. (C) 2003 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:136 / 141
页数:6
相关论文
共 26 条
[1]   Biases in a randomized comparison of three types of screw fixation in displaced femoral neck fractures [J].
Alho, A ;
Austdal, S ;
Benterud, JG ;
Blikra, G ;
Lerud, P ;
Raugstad, TS .
ACTA ORTHOPAEDICA SCANDINAVICA, 1998, 69 (05) :463-468
[2]  
BACKMAN S, 1957, PROXIMAL END FEMUR, P146
[3]  
BARNS R, 1981, J BONE JOINT SURG BR, V63, P357
[4]   HIP-JOINT LOADING DURING WALKING AND RUNNING, MEASURED IN 2 PATIENTS [J].
BERGMANN, G ;
GRAICHEN, F ;
ROHLMANN, A .
JOURNAL OF BIOMECHANICS, 1993, 26 (08) :969-990
[5]  
BOYCE J, 1986, J BONE JOINT SURG BR, V68, P156
[6]   FEMORAL-NECK FRACTURE FIXATION - COMPARISON OF A SLIDING SCREW WITH LAG SCREWS [J].
CLARK, DI ;
CROFTS, CE ;
SALEH, M .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1990, 72 (05) :797-800
[7]   STABILITY OF FIXATION IN FEMORAL-NECK FRACTURES - COMPARISON OF 4 FIXATION DEVICES INVIVO AND IN CADAVERS [J].
ELMERSON, S ;
ANDERSSON, GBJ ;
POPE, MH ;
ZETTERBERG, C .
ACTA ORTHOPAEDICA SCANDINAVICA, 1987, 58 (02) :109-112
[8]   THE STRUCTURE AND FUNCTION OF THE PROXIMAL END OF THE FEMUR [J].
GARDEN, RS .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1961, 43 (03) :576-589
[9]  
HEDLUND R, 1987, CLIN ORTHOP RELAT R, V222, P132
[10]  
Klenerman L, 1970, J Bone Joint Surg Br, V52, P514