Fast, sensitive and cost-effective detection of nerve agents in the gas phase using a portable instrument and an electrochemical biosensor

被引:71
作者
Arduini, Fabiana
Amine, Aziz
Moscone, Danila
Ricci, Francesco
Palleschi, Giuseppe
机构
[1] Univ Roma Tor Vergata, Dipartimento Sci & Technol Chim, I-00133 Rome, Italy
[2] Fac Sci & Tech, Mohammadia 20800, Morocco
关键词
biosensor; nerve agents; butyrylcholinesterase; screen printed electrode;
D O I
10.1007/s00216-007-1330-z
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The nerve agents are chemical warfare agents known to be used during terrorist attacks. An inexpensive and portable system to be used by first responders and military personnel is of interest owing to the continuing threat of possible terrorist attacks. Amperometric biosensors based on cholinesterase inhibition show such potentialities. In this work butyrylcholinesterase was immobilized onto screen-printed electrodes modified with Prussian blue and the nerve agent detection was performed by measuring the residual activity of enzyme. The optimized biosensor was tested with sarin and VX standard solutions, showing detection limits of 12 and 14 ppb (10% of inhibition), respectively. The enzymatic inhibition was also obtained by exposing the biosensors to sarin in gas phase. Two different concentrations of sarin gas (0.1 and 0.5 mg m(-3)) at different incubation times (from 30 s up to 10 min) were tested. It is possible to detect sarin at a concentration of 0.1 mg m(-3) with 30-s incubation time, with a degree of inhibition of 34%, which match the legal limits (immediate danger to life and health).
引用
收藏
页码:1049 / 1057
页数:9
相关论文
共 36 条
[31]   Determination of pesticides using electrochemical biosensors [J].
Trojanowicz, M ;
Hitchman, ML .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1996, 15 (01) :38-45
[32]  
TURNER APF, 1987, BIOSENSOR FUNDAMENTA
[33]   Toxicokinetics of the nerve agent (±)-VX in anesthetized and atropinized hairless guinea pigs and marmosets after intravenous and percutaneous administration [J].
van der Schans, MJ ;
Lander, BJ ;
van der Wiel, H ;
Langenberg, JP ;
Benschop, HP .
TOXICOLOGY AND APPLIED PHARMACOLOGY, 2003, 191 (01) :48-62
[34]   PULSED ELECTROCHEMICAL DETECTION OF CYSTEINE, CYSTINE, METHIONINE, AND GLUTATHIONE AT GOLD ELECTRODES FOLLOWING THEIR SEPARATION BY LIQUID-CHROMATOGRAPHY [J].
VANDEBERG, PJ ;
JOHNSON, DC .
ANALYTICAL CHEMISTRY, 1993, 65 (20) :2713-2718
[35]   A fluorescence-based biosensor for the detection of organophosphate pesticides and chemical warfare agents [J].
Viveros, L ;
Paliwal, S ;
McCrae, D ;
Wild, J ;
Simonian, A .
SENSORS AND ACTUATORS B-CHEMICAL, 2006, 115 (01) :150-157
[36]   Enzyme-based detection of sarin (GB) using planar waveguide absorbance spectroscopy [J].
White, BJ ;
Harmon, HJ .
SENSOR LETTERS, 2005, 3 (01) :36-41