Self-expanding nitinol stents: material and design considerations

被引:402
作者
Stoeckel, D [1 ]
Pelton, A [1 ]
Duerig, T [1 ]
机构
[1] Nitinol Devices & Components, Fremont, CA 94539 USA
关键词
self-expanding Nitinol stents; nickel; titanium; shape memory; superelasticity;
D O I
10.1007/s00330-003-2022-5
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Nitinol (nickel-titanium) alloys exhibit a combination of properties which make these alloys particularly suited for self-expanding stents. Some of these properties cannot be found in engineering materials used for stents presently. This article explains the fundamental mechanism of shape memory and superelasticity, and how they relate to the characteristic performance of self-expanding stents. Nitinol stents are manufactured to a size slightly larger than the target vessel size and delivered constrained in a delivery system. After deployment, they position themselves against the vessel wall with a low, 'chronic' outward force. They resist outside forces with a significantly higher radial resistive force. Despite the high nickel content of Nitinol, its corrosion resistance and biocompatibility is equal to that of other implant materials. The most common Nitinol stents are listed and described.
引用
收藏
页码:292 / 301
页数:10
相关论文
共 28 条
[1]  
*ASTM, 2002, F208201 ASTM
[2]  
*ASTM, 2002, F200400 ASTM
[3]  
*ASTM, 2002, F212901 ASTM
[4]  
*ASTM, 2002, F206300 ASTM
[5]   BIODEGRADATION OF ORTHODONTIC APPLIANCES .1. BIODEGRADATION OF NICKEL AND CHROMIUM INVITRO [J].
BARRETT, RD ;
BISHARA, SE ;
QUINN, JK .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 1993, 103 (01) :8-14
[6]   BIODEGRADATION OF ORTHODONTIC APPLIANCES .2. CHANGES IN THE BLOOD LEVEL OF NICKEL [J].
BISHARA, SE ;
BARRETT, RD ;
SELIM, MI .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 1993, 103 (02) :115-119
[7]   INVITRO STUDIES OF FRETTING CORROSION OF ORTHOPEDIC MATERIALS [J].
BROWN, SA ;
HUGHES, PJ ;
MERRITT, K .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1988, 6 (04) :572-579
[8]   TRANS-LUMINAL EXPANDABLE NITINOL COIL STENT GRAFTING - PRELIMINARY-REPORT [J].
DOTTER, CT ;
BUSCHMANN, RW ;
MCKINNEY, MK ;
ROSCH, J .
RADIOLOGY, 1983, 147 (01) :259-260
[9]   Physical properties of endovascular stents: An experimental comparison [J].
Duda, SH ;
Wiskirchen, J ;
Tepe, G ;
Bitzer, M ;
Kaulich, TW ;
Stoeckel, D ;
Claussen, CD .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2000, 11 (05) :645-654
[10]  
Duerig T.W., 2013, Engineering aspects of shape memory alloys, V1st ed.