Transient acid effects in inductively coupled plasma optical emission spectrometry and inductively coupled plasma mass spectrometry

被引:49
作者
Stewart, II [1 ]
Olesik, JW [1 ]
机构
[1] Ohio State Univ, Dept Geol Sci, Lab Plasma Spectrochem Laser Spect & Mass Spect, Columbus, OH 43210 USA
关键词
acid effect; nitric acid; inductively coupled plasma optical emission spectrometry; inductively coupled plasma mass spectrometry; matrix effects; aerosol properties; transport rates;
D O I
10.1039/a802236d
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A change in sample acid concentration affects both transient lover several minutes) and steady state ICP-OES and ICP-MS signal magnitudes, Steady state signals are depressed when the acid concentration is increased and vice versa. The transient effect appears as a signal 'undershoot' or 'overshoot' of up to 50% followed by a time dependent (5-25 min) 'relaxation' to the steady state signal, The magnitude of the transient acid effect depends on the type and concentration of acid, Time-resolved side-on (radial) ICP-OES, end-on (axial) ICP-OES, ICP-MS, aerosol size and transport rate measurements mere made. Similar acid effects mere observed in both ICP-OES and ICP-RIS, Following a change in sample acid concentration, the tertiary aerosol volumetric flux changed on a similar time scale as ICP-OES and ICP-MS signals whereas the primary aerosol flux did not. It is proposed that the transient effect is due to changes in the extent of aerosol evaporation in the spray chamber, which in turn affects the analyte transport rate, There appears to be a competition between evaporation of solution on the walls of the spray chamber and evaporation of the sample aerosol, The relative rates of aerosol and solution evaporation depend on their acid concentrations. The mater vapour pressure decreases as the acid concentration increases, Spray chambers conditioned with high acid concentrations will produce smaller amounts of water vapour from the solution on the spray chamber walls. When a low acid concentration sample is then sprayed into the spray chamber the aerosol evaporation mill be more extensive. With time, however, the acid concentration of the solution coating the malls decreases owing to dilution by the aerosol of lower acid concentration. Then the extent of evaporation of newly introduced aerosol mill decrease until a steady state is reached (i.e., the acid concentration of solution coating the walls equals the acid concentration in the aerosol), Transport rate measurements show an increase of 10% immediately after the sample is changed from a 25% to a 2% HNO3 matrix, Methods to minimize and potentially eliminate this effect are proposed.
引用
收藏
页码:843 / 854
页数:12
相关论文
共 50 条
[1]   PHASE DOPPLER SPRAY ANALYZER FOR SIMULTANEOUS MEASUREMENTS OF DROP SIZE AND VELOCITY DISTRIBUTIONS [J].
BACHALO, WD ;
HOUSER, MJ .
OPTICAL ENGINEERING, 1984, 23 (05) :583-590
[2]  
Bates L. C., 1991, THESIS U N CAROLINA
[3]   STUDY OF THE EFFECTS OF CONCOMITANT ELEMENTS IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
BEAUCHEMIN, D ;
MCLAREN, JW ;
BERMAN, SS .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1987, 42 (03) :467-490
[4]   REMOVAL OF A MATRIX EFFECT IN ICP-AES MULTIELEMENT ANALYSIS BY SIMPLEX OPTIMIZATION [J].
BELCHAMBER, RM ;
BETTERIDGE, D ;
WADE, AP ;
CRUICKSHANK, AJ ;
DAVISON, P .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1986, 41 (05) :503-505
[6]   EFFECT OF NITRIC AND HYDROCHLORIC ACIDS ON RARE-EARTH ELEMENT (REE) INTENSITIES IN INDUCTIVELY-COUPLED PLASMA EMISSION-SPECTROMETRY [J].
BRENNER, IB ;
MERMET, JM ;
SEGAL, I ;
LONG, GL .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1995, 50 (4-7) :323-331
[7]   STUDY OF THE DEPRESSIVE EFFECTS OF NITRIC-ACID ON THE LINE-INTENSITIES OF RARE-EARTH ELEMENTS IN INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY [J].
BRENNER, IB ;
SEGAL, I ;
MERMET, M ;
MERMET, JM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1995, 50 (4-7) :333-340
[8]   EFFECT OF ANALYTE AND SOLVENT TRANSPORT ON SIGNAL INTENSITIES IN INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
BROWNER, RF ;
CANALS, A ;
HERNANDIS, V .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1992, 47 (05) :659-673
[9]   FUNDAMENTAL-STUDIES ON PNEUMATIC GENERATION AND AEROSOL TRANSPORT IN ATOMIC SPECTROMETRY - EFFECT OF MINERAL ACIDS ON EMISSION INTENSITY IN INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY [J].
CANALS, A ;
HERNANDIS, V ;
TODOLI, JL ;
BROWNER, RF .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1995, 50 (4-7) :305-321
[10]   INFLUENCE OF THE SAMPLE INTRODUCTION SYSTEM ON ACID EFFECTS IN INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY [J].
CARRE, M ;
LEBAS, K ;
MARICHY, M ;
MERMET, M ;
POUSSEL, E ;
MERMET, JM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1995, 50 (4-7) :271-283