Bipolar and tetrapolar transfer impedance measurements from volume conductor

被引:38
作者
Brown, BH [1 ]
Wilson, AJ
Bertemes, P
机构
[1] Univ Sheffield, Royal Hallamshire Hosp, Dept Med Phys, Sheffield S10 2JF, S Yorkshire, England
[2] Univ Warwick, Dept Phys, Coventry CV4 7AL, W Midlands, England
关键词
D O I
10.1049/el:20001439
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Bipolar and tetrapolar impedance measurements from a volume of biological tissue are compared. In the bipolar case, sensitivity is always positive and decreases with depth. However, for tetrapolar measurements the mean sensitivity is found to be zero over the surface plane, rising to a maximum at a depth proportional to electrode spacing.
引用
收藏
页码:2060 / 2062
页数:3