Posterior motion preserving implants evaluated by means of intervertebral disc bulging and annular fiber strains

被引:18
作者
Heuer, Frank [1 ]
Schmidt, Hendrik [1 ]
Kaefer, Wolfram [2 ]
Graf, Nicolas [1 ]
Wilke, Hans-Joachim [1 ]
机构
[1] Univ Ulm, Inst Orthopaed Res & Biomech, D-89081 Ulm, Germany
[2] Univ Ulm, Dept Orthopaed Surg, D-89081 Ulm, Germany
关键词
Disc bulging; Posterior stabilization; Lumbar spine; Soft tissue deformation; Motion preserving implants laser scanner; SHEAR STRAIN; LUMBAR SPINE; PRESSURE; DEGENERATION; COMPRESSION; SYSTEM;
D O I
10.1016/j.clinbiomech.2011.09.004
中图分类号
R318 [生物医学工程];
学科分类号
100103 [病原生物学];
摘要
Background: The aims of motion preserving implants are to ensure sufficient stability to the spine, to release facet joints by also allowing a physiological loading to the intervertebral disc. The aim of this study was to assess disc load contribution by means of annular fiber strains and disc bulging of intact and stiffened segments. This was compared to the segments treated with various motion preserving implants. Methods: A laser scanning device was used to obtain three-dimensional disc bulging and annular fiber strains of six lumbar intervertebral discs (L2-3). Specimens were loaded with 500 Nor 7.5 Nm moments in a spine tester. Each specimen was treated with four different implants; DSS (TM), internal fixator. Coflex (TM), and TOPS (TM). Findings: In axial compression, all implants performed in a similar way. In flexion, the Coflex decreased range of motion by 13%, whereas bulging and fiber strains were similar to intact. The DSS stabilized segments by 54% compared to intact. TOPS showed a slight decrease in fiber strains (5%) with a range of motion similar to intact. The rigid fixator allowed strains up to 2%. In lateral bending, TOPS yielded range of motion values similar to intact, but maximum fiber strains doubled from 6.5% (intact) to 13.8%. Coflex showed range of motion, bulging and strain values similar to intact. The DSS and the rigid fixator reduced these values. The implants produced only minor changes in axial rotation. Interpretation: This study introduces an in vitro method, which was employed to evaluate spinal implants other than standard biomechanical methods. We could demonstrate that dynamic stabilization methods are able to keep fiber strains and disc bulging in a physiological range. (C) 2011 Elsevier Ltd. All rights reserved.
引用
收藏
页码:218 / 225
页数:8
相关论文
共 27 条
[1]
THE LUMBAR SPINE IN BACKWARD BENDING [J].
ADAMS, MA ;
DOLAN, P ;
HUTTON, WC .
SPINE, 1988, 13 (09) :1019-1026
[2]
Dynamic interspinous process technology [J].
Christie, SD ;
Song, JK ;
Fessler, RG .
SPINE, 2005, 30 (16) :S73-S78
[3]
Direct measurement of intervertebral disc maximum shear strain in six degrees of freedom: Motions that place disc tissue at risk of injury [J].
Costi, J. J. ;
Stokes, I. A. ;
Gardner-Morse, M. ;
Laible, J. P. ;
Scoffone, H. M. ;
Iatridis, J. C. .
JOURNAL OF BIOMECHANICS, 2007, 40 (11) :2457-2466
[4]
Tensile properties of nondegenerate human lumbar anulus fibrosus [J].
Ebara, S ;
Iatridis, JC ;
Setton, LA ;
Foster, RJ ;
Mow, VC ;
Weidenbaum, M .
SPINE, 1996, 21 (04) :452-461
[5]
Galante J.O., 1967, Acta Orthop Scand, V100, P101
[6]
A method to obtain surface strains of soft tissues using a laser scanning device [J].
Heuer, Frank ;
Wolfram, Uwe ;
Schmidt, Hendrik ;
Wilke, Hans-Joachim .
JOURNAL OF BIOMECHANICS, 2008, 41 (11) :2402-2410
[7]
The relation between intervertebral disc bulging and annular fiber associated strains for simple and complex loading [J].
Heuer, Frank ;
Schmidt, Hendrik ;
Wilke, Hans-Joachim .
JOURNAL OF BIOMECHANICS, 2008, 41 (05) :1086-1094
[8]
A new laser scanning technique for imaging intervertebral disc displacement and its application to modeling nucleotomy [J].
Heuer, Frank ;
Schmidt, Hendrik ;
Claes, Lutz ;
Wilke, Hans-Joachim .
CLINICAL BIOMECHANICS, 2008, 23 (03) :260-269
[9]
Creep associated changes in intervertebral disc bulging obtained with a laser scanning device [J].
Heuer, Frank ;
Schmitt, Herbert ;
Schmidt, Hendrik ;
Claes, Lutz ;
Wilke, Hans-Joachim .
CLINICAL BIOMECHANICS, 2007, 22 (07) :737-744
[10]
Single lamellar mechanics of the human lumbar anulus fibrosus [J].
Holzapfel, GA ;
Schulze-Bauer, CAJ ;
Feigl, G ;
Regitnig, P .
BIOMECHANICS AND MODELING IN MECHANOBIOLOGY, 2005, 3 (03) :125-140