Continuum x-ray source as a calibration system for charge coupled devices

被引:12
作者
Hashimotodani, K
Toneri, T
Kitamoto, S
Tsunemi, H
Hayashida, K
Miyata, E
Katayama, K
Kohomura, T
Asakura, R
Koyama, K
Yamamoto, K
Miyaguchi, K
Suzuki, H
机构
[1] Osaka Univ, Grad Sch Sci, Dept Earth & Space Sci, Toyonaka, Osaka 560, Japan
[2] Japan Sci & Technol Corp, CREST, Kawaguchi, Saitama 332, Japan
[3] Kyoto Univ, Grad Sch Sci, Dept Phys, Kyoto 60601, Japan
[4] Hamamatsu Photon KK, Hamamatsu, Shizuoka 435, Japan
关键词
D O I
10.1063/1.1148671
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We present x-ray spectra obtained by a charge coupled device (CCD) from an electron impacting x-ray source using several kinds of electron targets: Au, Al-bronze, Al, and pure Al. The x rays from the source are dispersed by a grating and the dispersed x rays are focused on the CCD. Owing to the fine spatial and moderate energy resolutions of the CCD, fine spectra with 2 eV energy resolution up to 2.2 keV are obtained. X rays from the purl Al target (99.999% purity) provide good continuum x rays except for Al K x rays. This continuum source is useful as a calibration source for low energy x-ray detectors. (C) 1998 American Institute of Physics.
引用
收藏
页码:392 / 395
页数:4
相关论文
共 6 条