First experiences with outpatient care of patients with implanted axial flow pumps

被引:43
作者
Wieselthaler, GM
Schima, H
Dworschak, M
Quittan, M
Nuhr, M
Czerny, M
Seebacher, G
Huber, L
Grimm, M
Wolner, E
机构
[1] Univ Vienna, Dept Cardiothorac Surg, Vienna, Austria
[2] Univ Vienna, LBI Cardiosurg Res, Vienna, Austria
[3] Univ Vienna, Dept Cardiac Anesthesiol, Vienna, Austria
[4] Univ Vienna, Dept Phys Med & Rehabil, Vienna, Austria
关键词
MicroMed-DeBakey ventricular assist device monitoring; physical training; hospital discharge;
D O I
10.1046/j.1525-1594.2001.025005331.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
As known from patients with pulsatile ventricular assist devices (VADs), early mobilization, physical exercise, and return to normal life are essential for optimal recovery. Recently, implantable rotary pumps became available for extended left ventricular support as bridges to transplantation. Modified procedures are essential for patient training and hospital discharge. The MicroMed-DeBakey VAD was implanted in 10 patients with endstage heart disease. After recovery, regular ergometer training was performed with loads adapted to the patient's condition. Procedures for patient observation under outdoor conditions and a blood pressure measuring device for low pulse pressure conditions were developed. Improvement of physical condition was achieved in 8 patients. In the first 2 patients, exercise capacity was limited due to flow obstruction. In the following patients, an increase of workload on the ergometer up to 120 W was observed. Correlated with training, lactate/load relationship and heart rate decreased. Three patients were discharged from the hospital during support. The DeBakey-VAD system can support patients for extended time periods and is suitable for recovery and exercise. Under optimal patient and environmental conditions, discharge from the hospital can be obtained.
引用
收藏
页码:331 / 335
页数:5
相关论文
共 10 条
[1]  
Allen GS, 1997, ARTIF ORGANS, V21, P922
[2]  
DEBAKEY ME, 1998, HEART REPLACEMENT AR, V6, P407
[3]   Novacor left ventricular assist system versus HeartMate vented electric left ventricular assist system as a long-term mechanical circulatory support device in bridging patients:: A prospective study [J].
El-Banayosy, A ;
Arusoglu, L ;
Kizner, L ;
Tenderich, G ;
Minami, K ;
Inoue, K ;
Körfer, R .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2000, 119 (03) :581-587
[4]   Mechanical circulatory support with the Novacor LVAS: World-wide clinical results [J].
Murali, S .
THORACIC AND CARDIOVASCULAR SURGEON, 1999, 47 :321-325
[5]   Diminished vasoconstrictive function caused by long-term nonpulsatile left heart bypass [J].
Nishimura, T ;
Tatsumi, E ;
Nishinaka, T ;
Taenaka, Y ;
Masuzawa, T ;
Nakata, M ;
Takano, H .
ARTIFICIAL ORGANS, 1999, 23 (08) :722-726
[6]   Cardiac autonomic nervous function during long-term nonpulsatile left heart bypass [J].
Nishinaka, T ;
Tatsumi, E ;
Nishimura, T ;
Taenaka, Y ;
Masuzawa, T ;
Nakata, M ;
Takano, H ;
Koyanagi, H .
ARTIFICIAL ORGANS, 1999, 23 (06) :500-503
[7]  
SCHIMA H, 2000, 8 M ISRP AACH, P34
[8]   First clinical experience with the DeBakey VAD continuous-axial-flow pump for bridge to transplantation [J].
Wieselthaler, GM ;
Schima, H ;
Hiesmayr, M ;
Pacher, R ;
Laufer, G ;
Noon, GP ;
DeBakey, M ;
Wolner, E .
CIRCULATION, 2000, 101 (04) :356-359
[9]  
WIESELTHALER GM, UNPUB SPECIAL CONSID
[10]  
WIESELTHALER GM, IN PRESS ANN THORAC