Selective comprehensive multi-dimensional separation for resolution enhancement in high performance liquid chromatography. Part I: Principles and instrumentation

被引:79
作者
Groskreutz, Stephen R. [1 ]
Swenson, Michael M. [1 ]
Secor, Laura B. [1 ]
Stoll, Dwight R. [1 ]
机构
[1] Gustavus Adolphus Coll, St Peter, MN 56082 USA
关键词
Two-dimensional; Multi-dimensional; HPLC; Focusing; Resolving power; ELECTRON-CAPTURE DETECTION; GAS-CHROMATOGRAPHY; PEAK-CAPACITY; DETERMINATION LIMITS; MASS-SPECTROMETRY; TRACE ANALYSIS; SAMPLES; ONLINE; GRADIENT; SOLVENT;
D O I
10.1016/j.chroma.2011.06.035
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
An approach to enhancing the resolution of select portions of conventional one-dimensional high performance liquid chromatography (HPLC) separations was developed, which we refer to as selective comprehensive two-dimensional HPLC (sLC x LC). In this first of a series of two papers we describe the principles of this approach, which breaks the long-standing link in on-line multi-dimensional chromatography between the timescales of sampling the first dimension (D-1) separation and the separation of fractions of D-1 effluent in the second dimension. This allows rapid, high-efficiency separations to be used in the first dimension, while still adequately sampling D-1 peaks. Transfer, transient storage, and subsequent second dimension (D-2) separations of multiple fractions of a particular D-1 peak produces a two-dimensional chromatogram that reveals the coordinates of the peak in both dimensions of the chromatographic space. Using existing valve technology we find that the approach is repeatable (%RSD of peak area <1.5%), even at very short first dimension sampling times - as low as 1 s. We have also systematically studied the critical influence of the volume and composition of fractions transferred from the first to the second dimension of the sLC x LC system with reversed-phase columns in both dimensions, and the second dimension operated isocratically. We find that dilution of the transferred fraction, so that it contains 10-20% less organic solvent than the D-2 eluent, generally mitigates the devastating effects. of large transfer volumes on D-2 performance in this type of system. Several example applications of the sLC x LC approach are described in the second part of this two-part series. We anticipate that future advances in the valve technology used here will significantly widen the scope of possible applications of the sLC x LC approach. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:31 / 40
页数:10
相关论文
共 50 条
[1]   AUTOMATED ONLINE MULTIDIMENSIONAL HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC TECHNIQUES FOR THE CLEANUP AND ANALYSIS OF WATER-SOLUBLE SAMPLES [J].
APFFEL, JA ;
ALFREDSON, TV ;
MAJORS, RE .
JOURNAL OF CHROMATOGRAPHY, 1981, 206 (01) :43-57
[2]   Accumulating resampling (modulation) in comprehensive two-dimensional capillary GC (GC x GC) [J].
Blumberg, Leonid M. .
JOURNAL OF SEPARATION SCIENCE, 2008, 31 (19) :3358-3365
[3]  
Carr P.W., 2011, ADV CHROMATOGRAPHY
[4]   Perspectives on Recent Advances in the Speed of High-Performance Liquid Chromatography [J].
Carr, Peter W. ;
Stoll, Dwight R. ;
Wang, Xiaoli .
ANALYTICAL CHEMISTRY, 2011, 83 (06) :1890-1900
[5]   A COMPARATIVE-STUDY OF LARGE VOLUME INJECTION TECHNIQUES FOR MICROBORE COLUMNS IN HPLC [J].
CLAESSENS, HA ;
KUYKEN, MAJ .
CHROMATOGRAPHIA, 1987, 23 (05) :331-336
[6]   Effect of first-dimension undersampling on effective peak capacity in comprehensive two-dimensional separations [J].
Davis, Joe M. ;
Stoll, Dwight R. ;
Carr, Peter W. .
ANALYTICAL CHEMISTRY, 2008, 80 (02) :461-473
[7]   Multidimensional liquid chromatographic separations applied to the analysis of food samples [J].
Dugo, P. ;
Kumm, T. ;
Cacciola, F. ;
Dugo, G. ;
Mondello, L. .
JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES, 2008, 31 (11-12) :1758-1807
[8]   MULTIDIMENSIONAL GAS-CHROMATOGRAPHY WITH ELECTRON-CAPTURE DETECTION FOR THE DETERMINATION OF TOXIC CONGENERS IN POLYCHLORINATED BIPHENYL MIXTURES [J].
DUINKER, JC ;
SCHULZ, DE ;
PETRICK, G .
ANALYTICAL CHEMISTRY, 1988, 60 (05) :478-482
[9]   Optimizing the peak capacity per unit time in one-dimensional and off-line two-dimensional liquid chromatography for the separation of complex peptide samples [J].
Eeltink, Sebastiaan ;
Dolman, Sebastiaan ;
Swart, Remco ;
Ursem, Mario ;
Schoenmakers, Peter J. .
JOURNAL OF CHROMATOGRAPHY A, 2009, 1216 (44) :7368-7374
[10]   2-DIMENSIONAL COLUMN LIQUID-CHROMATOGRAPHIC TECHNIQUE FOR RESOLUTION OF COMPLEX-MIXTURES [J].
ERNI, F ;
FREI, RW .