A portable analyzer for pouch-actuated, immunoassay cassettes

被引:38
作者
Qiu, Xianbo [1 ]
Liu, Changchun [1 ]
Mauk, Michael G. [1 ]
Hart, Robert W. [1 ]
Chen, Dafeng [1 ]
Qiu, Jing [1 ]
Kientz, Terry [1 ]
Fiene, Jonathan [1 ]
Bau, Haim H. [1 ]
机构
[1] Univ Penn, Dept Mech Engn & Appl Mech, Philadelphia, PA 19104 USA
关键词
Point of care test; Immunoassay; Lateral flow (LF); Microfluidic cassette; Pouch; Portable analyzer; Processor; Linear actuator; MICROFLUIDIC CASSETTE; INFECTIOUS-DISEASE; ASSAY; AMPLIFICATION;
D O I
10.1016/j.snb.2011.08.012
中图分类号
O65 [分析化学];
学科分类号
070302 [分析化学];
摘要
A portable, small footprint, light, general purpose analyzer (processor) to control the flow in immunoassay cassettes and to facilitate the detection of test results is described. The durable analyzer accepts disposable cassettes that contain pouches and reaction chambers for various unit operations such as hydration of dry reagents, stirring, and incubation. The analyzer includes individually controlled, linear actuators to compress the pouches in the cassette, which facilitates the pumping and mixing of sample and reagents, and to close diaphragm-based valves for flow control. The same types of actuators are used to compress pouches and actuate valves. The analyzer also houses a compact OEM scanner/reader to excite fluorescence and detect emission from labels. The analyzer is hydraulically isolated from the cassette, reducing the possibility of cross-contamination. The analyzer facilitates programmable, automated execution of a sequence of operations such as pumping and valving in a timely fashion, reducing the level of expertise required from the operator and the possibility for errors. The analyzer's design is modular and expandable to accommodate cassettes of various complexities and additional functionalities. In this paper, the utility of the analyzer has been demonstrated with the execution of a simple, consecutive, lateral flow assay of a model biological system and the test results were detected with up converting phosphor labels that are excited at infrared frequencies and emit in the visible spectrum. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:1529 / 1535
页数:7
相关论文
共 24 条
[1]
Bonenberger J., 2006, IVD TECH MAY
[2]
An integrated, self-contained microfluidic cassette for isolation, amplification, and detection of nucleic acids [J].
Chen, Dafeng ;
Mauk, Michael ;
Qiu, Xianbo ;
Liu, Changchun ;
Kim, Jitae ;
Ramprasad, Sudhir ;
Ongagna, Serge ;
Abrams, William R. ;
Malamud, Daniel ;
Corstjens, Paul L. A. M. ;
Bau, Haim H. .
BIOMEDICAL MICRODEVICES, 2010, 12 (04) :705-719
[3]
A user-friendly, highly sensitive assay to detect the IFN-γ secretion by T cells [J].
Corstjens, Paul L. A. M. ;
Zuiderwijk, Michel ;
Tanke, Hans J. ;
van der Ploeg-van Schip, Jolien J. ;
Ottenhoff, Tom H. M. ;
Geluk, Annemiek .
CLINICAL BIOCHEMISTRY, 2008, 41 (06) :440-444
[4]
Rapid assay format for multiplex detection of humoral immune responses to infectious disease pathogens (HIV, HCV, and TB) [J].
Corstjens, Paul L. A. M. ;
Chen, Zongyuang ;
Zuiderwijk, Michel ;
Bau, Haim H. ;
Abrams, William R. ;
Malamud, Daniel ;
Niedbala, R. Sam ;
Tanke, Hans J. .
ORAL-BASED DIAGNOSTICS, 2007, 1098 :437-445
[5]
Development of a test strip reader for a lateral flow membrane-based immunochromatographic assay [J].
Kim, S ;
Park, JK .
BIOTECHNOLOGY AND BIOPROCESS ENGINEERING, 2004, 9 (02) :127-131
[6]
Aptamer-based sensor arrays for the detection and quantitation of proteins [J].
Kirby, R ;
Cho, EJ ;
Gehrke, B ;
Bayer, T ;
Park, YS ;
Neikirk, DP ;
McDevitt, JT ;
Ellington, AD .
ANALYTICAL CHEMISTRY, 2004, 76 (14) :4066-4075
[7]
Optical scanner for immunoassays with up-converting phosphorescent labels [J].
Li, Janice J. ;
Ouellette, Amy L. ;
Giovangrandi, Laurent ;
Cooper, David E. ;
Ricco, Antonio J. ;
Kovacs, Gregory T. A. .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2008, 55 (05) :1560-1571
[8]
Rapid and Sensitive Detection of Protein Biomarker Using a Portable Fluorescence Biosensor Based on Quantum Dots and a Lateral Flow Test Strip [J].
Li, Zhaohui ;
Wang, Ying ;
Wang, Jun ;
Tang, Zhiwen ;
Pounds, Joel G. ;
Lin, Yuehe .
ANALYTICAL CHEMISTRY, 2010, 82 (16) :7008-7014
[9]
A timer-actuated immunoassay cassette for detecting molecular markers in oral fluids [J].
Liu, Changchun ;
Qiu, Xianbo ;
Ongagna, Serge ;
Chen, Dafeng ;
Chen, Zongyuan ;
Abrams, William R. ;
Malamud, Daniel ;
Corstjens, Paul L. A. M. ;
Bau, Haim H. .
LAB ON A CHIP, 2009, 9 (06) :768-776
[10]
Development and evaluation of a rapid lateral flow immunochromatographic strip assay for screening 19-nortestosterone [J].
Liu, Liqiang ;
Peng, Chifang ;
Jin, Zhengyu ;
Xu, Chuanlai .
BIOMEDICAL CHROMATOGRAPHY, 2007, 21 (08) :861-866