Computer model to predict subsurface damage in tibial inserts of total knees

被引:51
作者
Sathasivam, S [1 ]
Walker, PS [1 ]
机构
[1] UCL, Royal Natl Orthopaed Hosp Trust, Sch Med, Ctr Biomed Engn, Stanmore HA7 4LP, Middx, England
关键词
D O I
10.1002/jor.1100160507
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Two designs of total knee replacements were analysed to determine how the geometry of their bearing surface would affect the susceptibility of their ultra high molecular weight polyethylene tibial inserts to delamination. Orientations of the femoral components on the tibial surfaces were calculated with use of rigid body analysis for discrete intervals during the stance phase of gait. For each successive orientation, finite element analysis was used to compress the components together to determine the stresses in the tibial inserts. A damage function analogous to strain energy density was defined to account for the accumulated amplitudes and frequencies of the maximum shear stress cycles and hence to predict fatigue failure, The damage function was applied to each polyethylene element in the tibial insert, and the highest value calculated for each design was its damage score. One knee had a damage score more than three times less than that of the other because of lower stresses and because the contact points moved in the medial-lateral as well as anterior-posterior directions during internal-external rotation. The femoral and tibial components of this knee had large outer frontal radii and close conformity in the frontal plane. We propose that this method, which accounts for the motions and stresses endured during walking, makes different predictions regarding the likelihood of dt lamination compared with the predictions made by conventional static compression tests performed when the knee is in a neutral position.
引用
收藏
页码:564 / 571
页数:8
相关论文
共 29 条
[1]  
ANDERSON DD, 1995, T ORTHOP RES SOC, V20, P762
[2]  
ANDRIACCHI TP, 1995, COMMUNICATION
[3]   THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT [J].
BARTEL, DL ;
BICKNELL, VL ;
WRIGHT, TM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :1041-1051
[4]   POLYETHYLENE WEAR IN UNICONDYLAR KNEE PROSTHESES - 106 RETRIEVED MARMOR, PCA, AND ST-GEORG TIBIAL COMPONENTS COMPARED [J].
BLUNN, GW ;
JOSHI, AB ;
LILLEY, PA ;
ENGELBRECHT, E ;
RYD, L ;
LIDGREN, L ;
HARDINGE, K ;
NIEDER, E ;
WALKER, PS .
ACTA ORTHOPAEDICA SCANDINAVICA, 1992, 63 (03) :247-255
[5]   Wear in retrieved condylar knee arthroplasties - A comparison of wear in different designs of 280 retrieved condylar knee prostheses [J].
Blunn, GW ;
Joshi, AB ;
Minns, RJ ;
Lidgren, L ;
Lilley, P ;
Ryd, L ;
Engelbrecht, E ;
Walker, PS .
JOURNAL OF ARTHROPLASTY, 1997, 12 (03) :281-290
[6]  
BLUNN GW, 1996, T ORTHOP RES SOC, V21, P482
[7]  
BOSTROM MP, 1994, CLIN ORTHOP RELAT R, P20
[8]   Non-uniformity of contact stress on polyethylene inserts in total knee arthroplasty [J].
Bristol, RE ;
Fitzpatrick, DC ;
Brown, TD ;
Callaghan, JJ .
CLINICAL BIOMECHANICS, 1996, 11 (02) :75-80
[9]  
COLLIER JP, 1991, CLIN ORTHOP RELAT R, P232
[10]  
Ellyin F., 1989, BIAXIAL MULTIAXIAL F, P571