Automatic stepwise potentiometric titration in a monosegmented flow system

被引:16
作者
Borges, EP
Martelli, PB
Reis, BF
机构
[1] Univ Sao Paulo, Ctr Energia Nucl Agr, BR-13400970 Piracicaba, SP, Brazil
[2] Univ Sao Paulo, Inst Quim Sao Carlos, Sao Carlos, SP, Brazil
关键词
potentiometric titration; tubular pH electrode; step-wise titration; monosegmented flow; multicommutation;
D O I
10.1007/s006040070008
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An automatic stepwise potentiometric titration carried out in a monosegmented flow system is proposed. A tubular hydrogen ion-selective electrode without inner reference solution was employed as sensor. The titration procedure was implemented by exploiting continuous variations in volumetric fractions achieved by inserting different sample and titrant aliquots into the analytical path and maintaining the sample zone volume. Every sample plug was separated from the carrier solution by air bubbles to minimize dispersion and to distinguish each sample plug signal. The flow set-up comprised a set of computer-controlled three-way solenoid valves, and the software was written in Quick BASIC 4.5 allowing end-point determination while the titration was in progress. The feasibility of the system was ascertained by analyzing soft drinks, isotonic beverages, alcohol free beers and industrial fruit juices. Results were in agreement with those obtained by a conventional titration, and no significant difference at 95% confidence level was observed. Other profitable features such as a 1%. standard deviation (n = 10), and an analytical throughput of 22 samples per hour were also achieved.
引用
收藏
页码:179 / 184
页数:6
相关论文
共 22 条
[1]   SEQUENTIAL DETERMINATION OF CALCIUM AND NITRATE IONS IN WATERS BY POTENTIOMETRIC FLOW-INJECTION [J].
ALONSOCHAMARRO, J ;
BARTROLI, J ;
JUN, S ;
LIMA, JLFC ;
CONCEICAO, M ;
MONTENEGRO, BSM .
ANALYST, 1993, 118 (12) :1527-1532
[2]   CONTINUOUS AUTOMATED BURETLESS TITRATOR WITH DIRECT READOUT [J].
BLAEDEL, WJ ;
LAESSIG, RH .
ANALYTICAL CHEMISTRY, 1964, 36 (08) :1617-&
[3]  
COVINGTON KA, 1984, ION SELECTIVE ELECT, V1, P49
[4]   A flow-batch titrator exploiting a one-dimensional optimisation algorithm for end point search [J].
Honorato, RS ;
Araújo, MCU ;
Lima, RAC ;
Zagatto, EAG ;
Lapa, RAS ;
Lima, JLFC .
ANALYTICA CHIMICA ACTA, 1999, 396 (01) :91-97
[5]   BINARY SEARCH IN FLOW TITRATION EMPLOYING PHOTOMETRIC END-POINT DETECTION [J].
KORN, M ;
GOUVEIA, LFBP ;
DEOLIVEIRA, E ;
REIS, BF .
ANALYTICA CHIMICA ACTA, 1995, 313 (03) :177-184
[6]   AUTOMATED FLOW-INJECTION PSEUDOTITRATION OF STRONG AND WEAK ACIDS, ASCORBIC-ACID AND CALCIUM, AND CATALYTIC PSEUDOTITRATIONS OF AMINOPOLYCARBOXYLIC ACIDS BY USE OF A MICROCOMPUTER-CONTROLLED ANALYZER [J].
KOUPPARIS, MA ;
ANAGNOSTOPOULOU, P ;
MALMSTADT, HV .
TALANTA, 1985, 32 (05) :411-417
[7]   Automatic calibration in continuous flow analysis [J].
LopezGarcia, I ;
Vinas, P ;
HernandezCordoba, M .
ANALYTICA CHIMICA ACTA, 1996, 327 (02) :83-93
[8]   AUTOMATIC TITRATIONS IN UNSEGMENTED FLOW SYSTEMS BASED ON VARIABLE FLOW-RATE PATTERNS .2. COMPLEXOMETRIC AND REDOX TITRATIONS [J].
MARCOS, J ;
RIOS, A ;
VALCARCEL, M .
ANALYTICA CHIMICA ACTA, 1992, 261 (1-2) :495-503
[9]   Automatic potentiometric titration in monosegmented flow system exploiting binary search [J].
Martelli, PB ;
Reis, BF ;
Korn, M ;
Lima, JLFC .
ANALYTICA CHIMICA ACTA, 1999, 387 (02) :165-173
[10]  
Martelli PB, 1998, QUIM NOVA, V21, P133