The extended range neutron rem counter 'LINUS': Overview and latest developments

被引:97
作者
Birattari, C
Esposito, A
Ferrari, A
Pelliccioni, M
Rancati, T
Silari, M
机构
[1] Univ Milan, Dipartimento Fis, LASA, I-20090 Segrate, Italy
[2] Ist Nazl Fis Nucl, Lab Nazl Frascati, I-00044 Frascati, Italy
[3] Ist Nazl Fis Nucl, Sez Milano, I-20133 Milan, Italy
[4] CERN, CH-1211 Geneva 23, Switzerland
关键词
D O I
10.1093/oxfordjournals.rpd.a032258
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The 'history' of the extended range neutron rem counter LINUS, since its first conception in 1990, is reviewed, along with the latest developments. These include the calibration of the initially cylindrical version with nearly monoenergetic neutrons in the energy range 34-66 MeV, a detailed evaluation of the anisotropy of its response function, the construction and calibration of an improved spherical version, and recent measurements in reference high energy stray radiation fields. The instrument can now be considered as being fully characterised. Similar monitors built by other laboratories following the present design have confirmed its performance. The instrument is now in semi-routine use at a number of particle accelerator facilities and is one of several devices used on-board aircraft for assessing the exposure to cosmic rays at commercial flight altitudes.
引用
收藏
页码:135 / 148
页数:14
相关论文
共 40 条
[1]  
ANDERSSON IO, 1963, P IAEA S NEUTR DOS V, V2, P87
[2]  
[Anonymous], ICRP PUBL
[3]  
Ban S, 1996, PROCEEDINGS OF SECOND SPECIALISTS' MEETING ON SHIELDING ASPECTS OF ACCELERATORS, TARGETS AND IRRADIATION FACILITIES, P209
[4]   MEASUREMENTS AND CHARACTERIZATION OF HIGH-ENERGY NEUTRON FIELDS [J].
BIRATTARI, C ;
DEPONTI, E ;
ESPOSITO, A ;
FERRARI, A ;
PELLICCIONI, M ;
SILARI, M .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 338 (2-3) :534-543
[5]  
BIRATTARI C, 1992, RADIAT PROT DOSIM, V44, P193
[6]   AN EXTENDED RANGE NEUTRON REM COUNTER [J].
BIRATTARI, C ;
FERRARI, A ;
NUCCETELLI, C ;
PELLICCIONI, M ;
SILARI, M .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 297 (1-2) :250-257
[7]   CALIBRATION OF THE NEUTRON REM COUNTER LINUS IN THE ENERGY-RANGE FROM THERMAL TO 19 MEV [J].
BIRATTARI, C ;
ESPOSITO, A ;
FERRARI, A ;
PELLICCIONI, M ;
SILARI, M .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1993, 324 (1-2) :232-238
[8]   INTERCOMPARISON OF THE RESPONSE OF DOSIMETERS USED IN HIGH-ENERGY STRAY RADIATION-FIELDS [J].
BIRATTARI, C ;
ESPOSITO, A ;
FASSO, A ;
FERRARI, A ;
FESTAG, JG ;
HOFERT, M ;
NIELSEN, M ;
PELLICCIONI, M ;
RAFFNSOE, C ;
SCHMIDT, P ;
SILARI, M .
RADIATION PROTECTION DOSIMETRY, 1994, 51 (02) :87-94
[9]  
Birattari C, 1996, PROCEEDINGS OF SECOND SPECIALISTS' MEETING ON SHIELDING ASPECTS OF ACCELERATORS, TARGETS AND IRRADIATION FACILITIES, P171
[10]   SCANNING PROCEDURES FOR CALIBRATION OF RADIATION DETECTORS [J].
BIRATTARI, C ;
MANFREDI, G ;
SILARI, M .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (08) :4198-4203