Aliphatic polyurethane as a matrix for pH sensors: Effects of native sites and added proton carrier on electrical and potentiometric properties

被引:37
作者
Cosofret, VV
Erdosy, M
Raleigh, JS
Johnson, TA
Neuman, MR
Buck, RP
机构
[1] UNIV N CAROLINA,DEPT CHEM,CHAPEL HILL,NC 27599
[2] UNIV N CAROLINA,DEPT MED,CHAPEL HILL,NC 27599
[3] CASE WESTERN RESERVE UNIV,METROHLTH MED CTR,CLEVELAND,OH 44109
关键词
polyurethane; pH sensors; electrical properties; potentiometric properties;
D O I
10.1016/0039-9140(95)01724-0
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The potentiometric and impedance characteristics of polymeric membranes, based on aliphatic polyurethane (Tecoflex) as a matrix, are described and interpreted by theory and experiments for H+ and alkali metal ion-sensitive sensors. Both dummy plasticized membranes and proton carrier-loaded membranes can show pH response. The pH response of dummy membranes is due to protonated natural negative sites in the polyurethane matrix. The electrodes with added proton carrier show improved rejection of Li+, Na+, and K+ responses and give useful analytical responses. Optimal performance requires control of negative site concentration by addition of lipophilic salt (e.g. tetraphenylborate derivatives). Impedance analyses show surface-rate semicircles and, depending on the bathing electrolyte solution, appearance of a diffusional Warburg impedance. In addition to these time-dependence surface region effects, changes in the bulk membrane resistance with soaking time can be well correlated with equilibrium water content of plasticized membranes.
引用
收藏
页码:143 / 151
页数:9
相关论文
共 44 条
  • [1] [Anonymous], 1969, ANAL CHEM POLYURETHA
  • [2] ANZAI J, 1990, POLYM COMMUN, V31, P375
  • [3] BERGVELD P, 1988, WILSON WILSONS COMPR, V23
  • [4] BOUEN S, 1986, ANAL CHIM ACTA, V185, P347
  • [5] ANION EFFECTS ON DONNAN FAILURE OF AMINATED-POLY(VINYL CHLORIDE)-BASED AND NEUTRAL-CARRIER-BASED PH SENSORS
    BUCK, RP
    COSOFRET, VV
    LINDNER, E
    [J]. ANALYTICA CHIMICA ACTA, 1993, 282 (02) : 273 - 281
  • [6] BUCK RP, 1982, ION SEL ELECTRODE RE, V4, P3
  • [7] ELECTROCHEMICAL PERFORMANCE, BIOCOMPATIBILITY, AND ADHESION OF NEW POLYMER MATRICES FOR SOLID-STATE ION SENSORS
    CHA, GS
    LIU, D
    MEYERHOFF, ME
    CANTOR, HC
    MIDGLEY, AR
    GOLDBERG, HD
    BROWN, RB
    [J]. ANALYTICAL CHEMISTRY, 1991, 63 (17) : 1666 - 1672
  • [8] POTENTIOMETRIC ION-SELECTIVE AND BIO-SELECTIVE ELECTRODES BASED ON ASYMMETRIC CELLULOSE-ACETATE MEMBRANES
    CHA, GS
    MEYERHOFF, ME
    [J]. TALANTA, 1989, 36 (1-2) : 271 - 278
  • [9] POLYIMIDE-MATRIX CHEMICAL-SELECTIVE MEMBRANES
    CHA, GS
    BROWN, RB
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 1990, 1 (1-6) : 281 - 285
  • [10] EVIDENCE FOR A WATER-RICH SURFACE REGION IN POLY(VINYL CHLORIDE)-BASED ION-SELECTIVE ELECTRODE MEMBRANES
    CHAN, ADC
    LI, XZ
    HARRISON, DJ
    [J]. ANALYTICAL CHEMISTRY, 1992, 64 (21) : 2512 - 2517