Comparative study on the wear behaviour of different conventional and cross-linked polyethylenes for total hip replacement

被引:37
作者
Affatato, S. [1 ]
Zavalloni, M. [1 ]
Taddei, P. [2 ]
Di Foggia, M. [2 ]
Fagnano, C. [2 ]
Viceconti, M. [1 ]
机构
[1] Ist Ortoped Rizzoli, Lab Tecnol Med, I-40136 Bologna, Italy
[2] Univ Bologna, Sez Chim & Propedeut Biochim, Ctr Studio Spettroscopia Raman, Dipartimento Biochim G Moruzzi, Bologna, Italy
关键词
conventional ultra-high molecular weight polyethylene; cross-linked polyethylene; PE grade; sterilisation; hip joint simulator; micro-Raman spectroscopy;
D O I
10.1016/j.triboint.2008.02.006
中图分类号
TH [机械、仪表工业];
学科分类号
0802 [机械工程];
摘要
In this study, five different types of conventional and cross-linked polyethylene (XLPE) (gamma-sterilised PE GUR1020, EtO-sterilised PE GUR1020, gamma-sterilised PE GUR1050, EtO-sterilised XLPE GUR1020, EtO-sterilised XLPE GUR1050) acetabular cups were tested on a hip joint simulator run for 5 million cycles in order to compare the relative long-term wear resistance in relation to material properties (PE grade, conventional or cross-linked) and sterilisation method (EtO treatment or gamma-irradiation). Gravimetric measurements revealed significant differences between the wear behaviours of the five sets of acetabular cups. Weight loss was found to decrease along the series: gamma-sterilised PE GUR1020>EtO-sterilised PE GUR1020>gamma-sterilised PE GUR1050>EtO-sterilised XLPE GUR1050>EtO-sterilised XLPE GUR1020. The wear results were discussed in relation to the crystallinity degree of the cups which was determined by micro-Raman spectroscopy coupled to partial least-squares analysis. Within both conventional and crosslinked PE series, it appeared that higher crystallinity samples (i.e. gamma-sterilised PE GUR1020 and EtO-sterilised XLPE GUR1050, respectively) were characterised by higher wear rates. The higher weight loss observed for PE GUR1020 was explained in relation to its lower molecular weight with respect to PE GUR1050. Raman analysis showed that wear testing did not significantly modify the crystallinity degree of any of the tested acetabular cups. The most worn cup, i.e. gamma-sterilised PE GUR1020, appeared the most homogeneously polished upon wear testing, as confirmed by the lowest standard deviation associated to the crystallinity value recorded in the centre of the cup. The results of this investigation have clearly shown a dramatic wear reduction in favour of the cross-linked polyethylene. (C) 2008 Elsevier Ltd. All rights reserved.
引用
收藏
页码:813 / 822
页数:10
相关论文
共 65 条
[1]
Wear behaviour of cross-linked polyethylene assessed in vitro under severe conditions [J].
Affatato, S ;
Bersaglia, G ;
Rocchi, M ;
Taddei, P ;
Fagnano, C ;
Toni, A .
BIOMATERIALS, 2005, 26 (16) :3259-3267
[2]
The performance of gamma- and EtO-sterilised UHMWPE acetabular cups tested under severe simulator conditions. Part 2: wear particle characteristics with isolation protocols [J].
Affatato, S ;
Bersaglia, G ;
Emiliani, D ;
Foltran, I ;
Taddei, P ;
Reggiani, M ;
Ferrieri, P ;
Toni, A .
BIOMATERIALS, 2003, 24 (22) :4045-4055
[3]
Effects of the sterilisation method on the wear of UHMWPE acetabular cups tested in a hip joint simulator [J].
Affatato, S ;
Bordini, B ;
Fagnano, C ;
Taddei, P ;
Tinti, A ;
Toni, A .
BIOMATERIALS, 2002, 23 (06) :1439-1446
[4]
The effect of gamma irradiation and shelf aging in air on the oxidation of ultra-high molecular weight polyethylene [J].
Al-Ma'adeed, MA ;
Al-Quadawi, IY ;
Madi, N ;
Al-Thani, NJ .
APPLIED SURFACE SCIENCE, 2006, 252 (09) :3316-3322
[5]
BAJARIA SH, 1998, TECH PAPER SER, V3, P40
[6]
BANKSTON AB, 1995, CLIN ORTHOP RELAT R, P37
[7]
THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT [J].
BARTEL, DL ;
BICKNELL, VL ;
WRIGHT, TM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :1041-1051
[8]
Micro-Raman spectroscopy for the crystallinity characterization of UHMWPE hip cups run on joint simulators [J].
Bertoluzza, A ;
Fagnano, C ;
Rossi, M ;
Tinti, A ;
Cacciari, GL .
JOURNAL OF MOLECULAR STRUCTURE, 2000, 521 :89-95
[9]
Ultra high molecular-weight polyethylene (UHMWPE) in total knee replacement: Fabrication, sterilisation and wear [J].
Blunn, G ;
del Preva, EMB ;
Costa, L ;
Fisher, J ;
Freeman, MAR .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2002, 84B (07) :946-949
[10]
Influence of packaging conditions on the properties of gamma-irradiated UHMWPE following accelerated ageing and shelf ageing [J].
Buchanan, FJ ;
Sim, B ;
Downes, S .
BIOMATERIALS, 1999, 20 (09) :823-837