The possibility of using sodalime glass as a TL material for high-dose measurements and in accident dosimetry has been investigated. The structure of the glow curves, including the number of peaks was found to be dose and annealing temperature dependent. However, a linear relationship was obtained between the glass response and dose in the dose range studied. Variation in the responses of the glass after repeated use was obtained to be 4%. Variation in batch sensitivity was also found to be not more than 5%. The fading rate of the glass response at room temperature is small. (C) 2003 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:407 / 410
页数:4
相关论文
共 8 条
[1]   SITE-DEPENDENT THERMOLUMINESCENCE OF COPPER(I) IONS IN SILICA GLASS [J].
CHAUDHURI, J ;
DEBNATH, R .
JOURNAL OF PHYSICS-CONDENSED MATTER, 1994, 6 (21) :3987-3992
[2]   The thermoluminescence of Toshiba FD-1 and FD-7 RPL glass [J].
Croft, S ;
Weaver, DR ;
Matthews, R .
RADIATION PROTECTION DOSIMETRY, 1996, 65 (1-4) :381-384
[3]  
ERKNOL AY, 1995, RADIAT PHYS CHEM, V46, P111
[4]  
HOROWITZ YS, 1986, NUCL INSTRUM METH A, V243, P207, DOI 10.1016/0168-9002(86)90843-0
[5]   Fiber-optic-coupled, laser heated thermoluminescence dosimeter for remote radiation sensing [J].
Huston, AL ;
Justus, BL ;
Johnson, TL .
APPLIED PHYSICS LETTERS, 1996, 68 (24) :3377-3379
[6]  
JUSTUS BL, 1995, APPL PHYS LETT, V67, P1779
[7]   ACCIDENT DOSIMETRY USING THE TL FROM DENTAL RESTORATION PORCELAINS [J].
MAURICIO, CLP ;
ROSA, LAR ;
CUNHA, PG .
RADIATION PROTECTION DOSIMETRY, 1985, 11 (03) :185-188
[8]  
MCKEEVER SWS, 1983, PACT, V9, P123