Disposable strip potentiometric electrodes with solvent-polymeric ion-selective membranes fabricated using screen-printing technology

被引:51
作者
Koncki, R
Glab, S
Dziwulska, J
Palchetti, I
Mascini, M
机构
[1] Univ Warsaw, Dept Chem, PL-02093 Warsaw, Poland
[2] Univ Florence, Dipartimento Sanita Pubbl, Sez Chim Analit, I-50121 Florence, Italy
关键词
potentiometry; ion-selective electrodes; thick-film sensors; screen-printing technology;
D O I
10.1016/S0003-2670(98)00726-0
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple and reproducible method for mass production of disposable, plastic, thick-film, ion-selective electrodes using screen-printing technology is described. The properties of selective sensors for potassium, ammonium and nitrate ions obtained with this technology are shown and discussed. All the obtained strip sensors have analytical characteristics (sensitivity, selectivity, dynamic range and response time) comparable with those for conventional ion-selective electrodes. However, they practically do not require any conditioning before measurements and they respond in short time (steady-state signal is reached after a few seconds), that is important in case of the disposable sensors. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:451 / 459
页数:9
相关论文
共 22 条
[21]   SYNTHESIS AND CHARACTERIZATION OF NEW POLYVINYLCHLORIDE MEMBRANES FOR ENHANCED ADHESION ON ELECTRODE SURFACES [J].
SATCHWILL, T ;
HARRISON, DJ .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1986, 202 (1-2) :75-81
[22]  
Umezawa Y., 1990, CRC HDB ION SELECTIV