Room shielding for intensity-modulated radiation therapy treatment facilities

被引:32
作者
Mutic, S [1 ]
Low, DA [1 ]
Klein, EE [1 ]
Dempsey, JF [1 ]
Purdy, JA [1 ]
机构
[1] Edward Mallinckrodt Inst Radiol, Radiat Oncol Ctr, St Louis, MO 63110 USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2001年 / 50卷 / 01期
关键词
shielding; IMRT; DMLC; serial tomotherapy; leakage;
D O I
10.1016/S0360-3016(01)01463-8
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: The traditional assumptions used in room-shielding calculations are reassessed for intensity-modulated radiation therapy (IMRT), IMRT makes relatively inefficient use of monitor units (MUs) when compared to conventional radiation therapy, affecting the assumptions used in room-shielding calculations, For the same single-fraction tumor dose delivered, the total number of MUs for IMRT is much greater than for a conventional treatment. Therefore, the exposure contribution from the linear accelerator head leakage will be significantly greater than with conventional treatments, Methods and Materials: We propose a shielding calculation model that decouples the concepts of workload, MUs, and target dose when determining primary and secondary barrier thicknesses. The workload for primary barrier calculations for conventional multileaf collimator (MLC) IMRT treatments is determined according to patient tumor doses, The same calculation for accelerator-based serial tomotherapy IMRT requires scaling by the average number of treatment slices. However, rotational therapy yields a small use factor that compensates for this increase, We further define a series of efficiency factors to account for the small field sizes employed in IMRT, For secondary barrier calculations, the patient-scattered radiation is assumed to be the same for all IMRT modalities as for conventional therapy. The accelerator head leakage contribution is proportional to the number of MUs, Knowledge of the average number of MUs per patient is required to estimate the head leakage contribution. We used a 6-MV linear accelerator photon beam to guide the development of this technique and to evaluate the adequacy of conventional barriers for IMRT, Average weekly IMRT workload estimates were made based on our experience with 180 serial tomotherapy patients and published data for both "step and shoot" and dynamic MLC delivered treatments, Results: We found that conventional primary barriers are adequate for both dynamic MLC and serial tomotherapy IMRT, However, the excessive head leakage produced by these modalities requires an increase in secondary barrier shielding. Conclusion: When designing shielding for an IMRT facility, increases in accelerator head leakage must he taken into account for secondary shielding, Adequacy of secondary shielding will depend on the IMRT patient load. For conventional facilities that are being assessed for IMRT therapy, existing primary barriers will typically prove adequate. (C) 2001 Elsevier Science Inc.
引用
收藏
页码:239 / 246
页数:8
相关论文
共 22 条
  • [1] Theoretical considerations of monitor unit calculations for intensity modulated beam treatment planning
    Boyer, A
    Xing, L
    Ma, CM
    Curran, B
    Hill, R
    Kania, A
    Bleier, A
    [J]. MEDICAL PHYSICS, 1999, 26 (02) : 187 - 195
  • [2] Modulated beam conformal therapy for head and neck tumors
    Boyer, AL
    Geis, P
    Grant, W
    Carol, M
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 39 (01): : 227 - 236
  • [3] Initial clinical experience with the Peacock intensity modulation of a 3-D conformal radiation therapy system
    Carol, M
    Grant, WH
    Pavord, D
    Eddy, P
    Targovnik, HS
    Butler, B
    Woo, S
    Figura, J
    Onufrey, V
    Grossman, R
    Selkar, R
    [J]. STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 1996, 66 (1-3) : 30 - 34
  • [4] PEACOCK(TM) - A SYSTEM FOR PLANNING AND ROTATIONAL DELIVERY OF INTENSITY-MODULATED FIELDS
    CAROL, MP
    [J]. INTERNATIONAL JOURNAL OF IMAGING SYSTEMS AND TECHNOLOGY, 1995, 6 (01) : 56 - 61
  • [5] THE GENERATION OF INTENSITY-MODULATED FIELDS FOR CONFORMAL RADIOTHERAPY BY DYNAMIC COLLIMATION
    CONVERY, DJ
    ROSENBLOOM, ME
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (06) : 1359 - 1374
  • [6] Estimates of whole-body dose equivalent produced by beam intensity modulated conformal therapy
    Followill, D
    Geis, P
    Boyer, A
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (03): : 667 - 672
  • [7] Grant W 3rd, 1998, Med Dosim, V23, P237, DOI 10.1016/S0958-3947(98)00015-6
  • [8] SIMULTANEOUS-OPTIMIZATION OF DYNAMIC MULTILEAF COLLIMATION AND SCANNING PATTERNS OR COMPENSATION FILTERS USING A GENERALIZED PENCIL BEAM ALGORITHM
    GUSTAFSSON, A
    LIND, BK
    SVENSSON, R
    BRAHME, A
    [J]. MEDICAL PHYSICS, 1995, 22 (07) : 1141 - 1156
  • [9] KAHN FM, 1994, PHYSICS RAD THERAPY
  • [10] Quantitative dosimetric verification of an IMRT planning and delivery system
    Low, DA
    Mutic, S
    Dempsey, JF
    Gerber, RL
    Bosch, WR
    Perez, CA
    Purdy, JA
    [J]. RADIOTHERAPY AND ONCOLOGY, 1998, 49 (03) : 305 - 316