Sterilization by gamma radiation impairs the tensile fatigue life of cortical bone by two orders of magnitude

被引:46
作者
Akkus, O [1 ]
Belaney, RM [1 ]
机构
[1] Univ Toledo, Dept Bioengn, Werribee, Vic, Australia
关键词
bone graft; gamma radiation; sterilization; cortical bone; fatigue;
D O I
10.1016/j.orthres.2005.03.003
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Cortical bone grafts are utilized frequently for skeletal reconstruction, spinal fusion and tumor surgery. Due to its efficacy and convenience terminal sterilization by gamma radiation is often essential to minimize disease transmission and infection. However, the impairment in the material properties of bone tissue secondary to gamma radiation sterilization is a concern since the mechanical functionality of a bone graft is of primary importance. While the extent of this impairment is well investigated for monotonic loading conditions, there does not seem to exist any information on the effects of gamma radiation sterilization on cortical bone's fatigue properties, the physiologically relevant mode of loading. In this study we investigated the degradation in the high-cycle and low-cycle tensile fatigue lives of cortical bone tissue secondary to gamma radiation sterilization at a dose of 36.4 kGy which approximately falls in the higher end of the standard dose range used in tissue banking. The high-cycle and the low-cycle fatigue tests were conducted under load control at initial strain levels of 0.2% and 0.4%, respectively. Monotonic tensile tests were also conducted to compare the impairment of fatigue properties with the impairment of monotonic properties, Results demonstrated that the impairment in both the high-cycle and the low-cycle fatigue lives were two orders of magnitude following sterilization, a change much more pronounced than that observed for monotonic loading. In conclusion, the results suggest that the impairment of the mechanical function of gamma radiation sterilized allografts is even worse in fatigue than monotonically. Therefore, grafts should be designed to minimize functional strains and avoid stress raisers to prevent premature fatigue failures. (c) 2005 Orthopaedic Research Society. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:1054 / 1058
页数:5
相关论文
共 26 条
[1]   Microstructural aspects of the fracture process in human cortical bone [J].
Akkus, O ;
Jepsen, KJ ;
Rimnac, CM .
JOURNAL OF MATERIALS SCIENCE, 2000, 35 (24) :6065-6074
[2]   Cortical bone tissue resists fatigue fracture by deceleration and arrest of microcrack growth [J].
Akkus, O ;
Rimnac, CM .
JOURNAL OF BIOMECHANICS, 2001, 34 (06) :757-764
[3]   Fracture resistance of gamma radiation sterilized cortical bone allografts [J].
Akkus, O ;
Rimnac, CM .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2001, 19 (05) :927-934
[4]   Allograft bone - The influence of processing on safety and performance [J].
Boyce, T ;
Edwards, J ;
Scarborough, N .
ORTHOPEDIC CLINICS OF NORTH AMERICA, 1999, 30 (04) :571-+
[5]  
Bright R. W., 1984, OSTEOCHONDRAL ALLOGR, P241
[6]  
BRIGHT RW, 1984, OSTEOCHONDRAL ALLOGR, P223
[7]  
Bruce Martin R., 1998, Skeletal Tissue Mechanics
[8]   Effect of freeze-drying and gamma irradiation on the mechanical properties of human cancellous bone [J].
Cornu, O ;
Banse, X ;
Docquier, PL ;
Luyckx, S ;
Delloye, C .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2000, 18 (03) :426-431
[9]   Effects of ionizing radiation on the mechanical properties of human bone [J].
Currey, JD ;
Foreman, J ;
Laketic, I ;
Mitchell, J ;
Pegg, DE ;
Reilly, GC .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1997, 15 (01) :111-117
[10]   Biomechanical issues in bone transplantation [J].
Davy, DT .
ORTHOPEDIC CLINICS OF NORTH AMERICA, 1999, 30 (04) :553-+