A new label-free amperometric immunosenor for rubella vaccine

被引:16
作者
Zhang, LY
Yuan, R [1 ]
Huang, XQ
Chai, YQ
Tang, DP
Cao, SR
机构
[1] SW China Normal Univ, Coll Chem & Chem Engn, Chongqing Key Lab Analyt Chem, Chongqing 400715, Peoples R China
[2] Tianjin Polytech Univ, Sch Mat Sci & Chem Engn, Tianjin 300160, Peoples R China
基金
中国国家自然科学基金;
关键词
label-free; amperometric; rubella vaccine; immunosensor; nano-Au;
D O I
10.1007/s00216-004-3021-3
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A novel label-free amperometric immunosensor for the detection of rubella vaccine was developed by immobilizing anti-rubella serum on bilayer nano-Au/polymerized o-phenylenediamine film with electrodeposited Prussian Blue (PB) as an electrode transfer mediator on the platinum electrode. The redox reactions of PB as a probe on the platinum surface were blocked due to the binding of the antibody to the antigen, which was investigated by cyclic voltammetry. Therefore, the interaction of the antibody with various concentrations of antigen could be detected by measurements of amperometric response in PBS, and the amperometric response on the surface of the modified electrode was inversely proportional to the concentration of rubella vaccine in the sample. The immunosensor showed a specific response to rubella vaccine in the range 8.1x10(-8) -8.0x10(-6) lgCCID(50)/ml (cell culture infectious dose) and a detection limit of 4.010(-8) lgCCID(50)/ml at a signal-to-noise ratio of 3. To summarize, the present work provides a low-cost, fast response time, highly sensitive and easy-to-prepare method for the determination of antigen in biological products.
引用
收藏
页码:1036 / 1040
页数:5
相关论文
共 31 条
[1]   Detection of antigen-antibody binding events with the atomic force microscope [J].
Allen, S ;
Chen, XY ;
Davies, J ;
Davies, MC ;
Dawkes, AC ;
Edwards, JC ;
Roberts, CJ ;
Sefton, J ;
Tendler, SJB ;
Williams, PM .
BIOCHEMISTRY, 1997, 36 (24) :7457-7463
[2]   Evaluation of antibody immobilization methods for piezoelectric biosensor application [J].
Babacan, S ;
Pivarnik, P ;
Letcher, S ;
Rand, AG .
BIOSENSORS & BIOELECTRONICS, 2000, 15 (11-12) :615-621
[3]   Capacitance measurements of antibody-antigen interactions in a flow system [J].
Berggren, C ;
Johansson, G .
ANALYTICAL CHEMISTRY, 1997, 69 (18) :3651-3657
[4]  
BERGGREN C, 1997, ANAL CHEM, V67, P1245
[5]   A NOVEL BIOSENSOR ELABORATION BY ELECTROPOLYMERIZATION OF AN ADSORBED AMPHIPHILIC PYRROLE TYROSINASE ENZYME LAYER [J].
COSNIER, S ;
INNOCENT, C .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1992, 328 (1-2) :361-366
[6]   Amperometric detection of pyridine nucleotides via immobilized viologen-accepting pyridine nucleotide oxidoreductase or immobilized diaphorase [J].
Cosnier, S ;
LeLous, K .
TALANTA, 1996, 43 (03) :331-337
[7]   SEPARATION-FREE SANDWICH ENZYME IMMUNOASSAYS USING MICROPOROUS GOLD ELECTRODES AND SELF-ASSEMBLED MONOLAYER IMMOBILIZED CAPTURE ANTIBODIES [J].
DUAN, CM ;
MEYERHOFF, ME .
ANALYTICAL CHEMISTRY, 1994, 66 (09) :1369-1377
[8]   Electric potential control of DNA immobilization on gold electrode [J].
Ge, CW ;
Liao, JH ;
Yu, W ;
Gu, N .
BIOSENSORS & BIOELECTRONICS, 2003, 18 (01) :53-58
[9]   SIMULTANEOUS IMMUNOASSAY USING ELECTROCHEMICAL DETECTION OF METAL-ION LABELS [J].
HAYES, FJ ;
HALSALL, HB ;
HEINEMAN, WR .
ANALYTICAL CHEMISTRY, 1994, 66 (11) :1860-1865
[10]   IMPROVED FIBER OPTIC CHEMICAL SENSOR FOR PENICILLIN [J].
HEALEY, BG ;
WALT, DR .
ANALYTICAL CHEMISTRY, 1995, 67 (24) :4471-4476