Permanent Pacemaker Insertion After CoreValve Transcatheter Aortic Valve Implantation Incidence and Contributing Factors (the UK CoreValve Collaborative)

被引:253
作者
Khawaja, M. Z. [1 ]
Rajani, R. [1 ]
Cook, A. [1 ]
Khavandi, A. [1 ]
Moynagh, A. [1 ]
Chowdhary, S. [1 ]
Spence, M. S. [1 ]
Brown, S. [1 ]
Khan, S. Q. [1 ]
Walker, N. [1 ]
Trivedi, U. [1 ]
Hutchinson, N. [1 ]
De Belder, A. J. [1 ]
Moat, N. [1 ]
Blackman, D. J. [1 ]
Levy, R. D. [1 ]
Manoharan, G. [1 ]
Roberts, D. [1 ]
Khogali, S. S. [1 ]
Crean, P. [1 ]
Brecker, S. J. [1 ]
Baumbach, A. [1 ]
Mullen, M. [1 ]
Laborde, J. -C. [1 ]
Hildick-Smith, D. [1 ]
机构
[1] Brighton & Sussex Univ Hosp Trust, Sussex Cardiac Ctr, UK CoreValve Collaborat, Brighton BN2 5BE, E Sussex, England
关键词
aortic stenosis; electrocardiography; pacemakers; transcatheter aortic valve; transcatheter aortic valve implantation; HIGH-RISK PATIENTS; VALVULAR HEART-DISEASE; PROGNOSTIC-SIGNIFICANCE; REPLACEMENT; STENOSIS; CONDUCTION; PROSTHESIS; BLOCK; ABNORMALITIES; REQUIREMENT;
D O I
10.1161/CIRCULATIONAHA.109.927152
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background-Permanent pacemaker (PPM) requirement is a recognized complication of transcatheter aortic valve implantation. We assessed the UK incidence of permanent pacing within 30 days of CoreValve implantation and formulated an anatomic and electrophysiological model. Methods and Results-Data from 270 patients at 10 centers in the United Kingdom were examined. Twenty-five patients (8%) had preexisting PPMs; 2 patients had incomplete data. The remaining 243 were 81.3 +/- 6.7 years of age; 50.6% were male. QRS duration increased from 105 +/- 23 to 135 +/- 29 milliseconds (P < 0.01). Left bundle-branch block incidence was 13% at baseline and 61% after the procedure (P < 0.001). Eighty-one patients (33.3%) required a PPM within 30 days. Rates of pacing according to preexisting ECG abnormalities were as follows: right bundle-branch block, 65.2%; left bundle-branch block, 43.75%; normal QRS, 27.6%. Among patients who required PPM implantation, the median time to insertion was 4.0 days (interquartile range, 2.0 to 7.75 days). Multivariable analysis revealed that periprocedural atrioventricular block (odds ratio, 6.29; 95% confidence interval, 3.55 to 11.15), balloon predilatation (odds ratio, 2.68; 95% confidence interval, 2.00 to 3.47), use of the larger (29 mm) CoreValve prosthesis (odds ratio, 2.50; 95% confidence interval, 1.22 to 5.11), interventricular septum diameter (odds ratio, 1.18; 95% confidence interval, 1.10 to 3.06), and prolonged QRS duration (odds ratio, 3.45; 95% confidence interval, 1.61 to 7.40) were independently associated with the need for PPM. Conclusion-One third of patients undergoing a CoreValve transcatheter aortic valve implantation procedure require a PPM within 30 days. Periprocedural atrioventricular block, balloon predilatation, use of the larger CoreValve prosthesis, increased interventricular septum diameter and prolonged QRS duration were associated with the need for PPM. (Circulation. 2011; 123: 951-960.)
引用
收藏
页码:951 / U47
页数:12
相关论文
共 42 条
[1]   The potential of myocardial perfusion scintigraphy for risk stratification of asymptomatic patients with type 2 diabetes [J].
Bax, Jeroen J. ;
Bonow, Robert O. ;
Tschoepe, Diethelm ;
Inzucchi, Silvio E. ;
Barrett, Eugene .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2006, 48 (04) :754-760
[2]   Percutaneous Aortic Valve Implants Under Sedation: Our Initial Experience [J].
Behan, Miles ;
Haworth, Peter ;
Hutchinson, Nevil ;
Trivedi, Uday ;
Laborde, Jean-Claude ;
Hildick-Smith, David .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2008, 72 (07) :1012-1015
[3]   Percutaneous transcatheter implantation of an aortic valve prosthesis for calcific aortic stenosis - First human case description [J].
Cribier, A ;
Eltchaninoff, H ;
Bash, A ;
Borenstein, N ;
Tron, C ;
Bauer, F ;
Derumeaux, G ;
Anselme, F ;
Laborde, F ;
Leon, MB .
CIRCULATION, 2002, 106 (24) :3006-3008
[4]   Treatment of calcific aortic stenosis with the percutaneous heart valve - Mid-term follow-up from the initial feasibility studies: The French experience [J].
Cribier, A ;
Eltchaninoff, H ;
Tron, C ;
Bauer, F ;
Agatiello, C ;
Nercolini, D ;
Tapiero, S ;
Litzler, PY ;
Bessou, JP ;
Babaliaros, V .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2006, 47 (06) :1214-1223
[5]   Permanent pacemaker implantation after isolated aortic valve replacement: Incidence, indications, and predictors [J].
Dawkins, Sam ;
Hobson, Alex R. ;
Kalra, Paul R. ;
Tang, Augustine T. M. ;
Monro, James L. ;
Dawkins, Keith D. .
ANNALS OF THORACIC SURGERY, 2008, 85 (01) :108-112
[6]   High-Risk Patients Referred for Transcatheter Aortic Valve Implantation: Management and Outcomes [J].
Dewey, Todd M. ;
Brown, David L. ;
Das, Tony S. ;
Ryan, William H. ;
Fowler, Jill E. ;
Hoffman, Shannon D. ;
Prince, Syma L. ;
Herbert, Morley A. ;
Culica, Dan ;
Mack, Michael J. .
ANNALS OF THORACIC SURGERY, 2008, 86 (05) :1450-1457
[7]   SITES OF CONDUCTION DISEASE IN AORTIC-STENOSIS - SIGNIFICANCE OF VALVE GRADIENT AND CALCIFICATION [J].
DHINGRA, RC ;
AMATYLEON, F ;
PIETRAS, RJ ;
WYNDHAM, C ;
DEEDWANIA, PC ;
WU, D ;
DENES, P ;
ROSEN, KM .
ANNALS OF INTERNAL MEDICINE, 1977, 87 (03) :275-280
[8]   A comparison of balloon- and self-expanding stents [J].
Duerig, TW ;
Wholey, M .
MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 2002, 11 (04) :173-178
[9]   Prognostic significance of newly acquired bundle branch block after aortic valve replacement [J].
El-Khally, Z ;
Thibault, B ;
Staniloae, C ;
Theroux, P ;
Dubuc, M ;
Roy, D ;
Guerra, P ;
Macle, L ;
Talajic, M .
AMERICAN JOURNAL OF CARDIOLOGY, 2004, 94 (08) :1008-1011
[10]   The bioprosthesis type and size influence the postoperative incidence of permanent pacemaker implantation in patients undergoing aortic valve surgery [J].
Elahi, Maqsood ;
Usmaan, Khalid .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2006, 15 (02) :113-118