Electromagnetic interference on medical equipment by low-power mobile telecommunication systems

被引:49
作者
Hanada, E [1 ]
Antoku, Y
Tani, S
Kimura, M
Hasegawa, A
Urano, S
Ohe, K
Yamaki, M
Nose, Y
机构
[1] Kyushu Univ, Grad Sch Med Sci, Dept Med Informat Sci, Fukuoka 8128582, Japan
[2] Hamamatsu Univ Sch Med, Dept Med Informat, Hamamatsu, Shizuoka 4313192, Japan
[3] Gunma Univ, Sch Med, Dept Internal Med 2, Maebashi, Gumma 3718511, Japan
[4] Niigata Univ Hosp, Operat Ctr, Niigata 9518520, Japan
[5] Univ Tokyo, Hosp Comp Ctr, Tokyo 1138655, Japan
[6] Yamagata Univ, Sch Med, Div Med Informat, Yamagata 9909585, Japan
关键词
electromagnetic (EM) field; electromagnetic interference (EMI); electronic medical equipment; mobile telecommunication systems; personal handy-phone system (PHS);
D O I
10.1109/15.902316
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
There have been a number of reports of electromagnetic interference (EMI) on electronic medical equipment caused by mobile telecommunication systems. In Japan, the use of the personal handy-phone system (PHS) has greatly expanded within urban areas. PHS handsets transmit EM signals at a frequency of 1.9 GHz and have a peak radiated power of 80 mW. This power level is lo,ver than that of other mobile telecommunication systems, Two studies were carried out. One was to determine whether or not PHS interferes with electronic medical equipment in hospitals. We observed no EMI on electronic medical equipment when the PHS handset was in either the speaking mode or on standby, The second study was to observe EMI from 1.9-GHz signals at several radiation power levels. Although EMI was not observed at the radiated peak power of the PHS handset, EMI on some of the tested equipment was observed when the radiated power was ten or more times higher than that of the PHS handset.
引用
收藏
页码:470 / 476
页数:7
相关论文
共 19 条
[1]  
*AAMI TIR, 1997, 18 AAMI TIR
[2]  
*AM NAT STAND I, 1997, C6318 ANSI
[3]   DO EUROPEAN GSM MOBILE CELLULAR PHONES POSE A POTENTIAL RISK TO PACEMAKER PATIENTS [J].
BARBARO, V ;
BARTOLINI, P ;
DONATO, A ;
MILITELLO, C ;
ALTAMURA, G ;
AMMIRATI, F ;
SANTINI, M .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (06) :1218-1224
[4]  
Clifford K J, 1994, Australas Phys Eng Sci Med, V17, P23
[5]  
Hamilton J, 1996, CAN MED ASSOC J, V154, P373
[6]  
*HLTH PROT BRANCH, 1995, DIG CELL PHON INT CA
[7]  
*INT EL COMM, 1995, 100043 IEC
[8]  
*INT EL COMM, 1993, 60112 IEC
[9]  
*MIN HLTH WELF PHA, 1996, INF ADV REACT DRUGS
[10]  
SILBERBERG JL, 1993, COMPLIANCE ENG FAL, P25