Validation of a new transit time ultrasound flowmeter in man

被引:94
作者
Laustsen, J
Pedersen, EM
Terp, K
Steinbruchel, D
Kure, HH
Paulsen, PK
Jorgensen, H
Paaske, WP
机构
[1] Dept. Cardiothoracic Vasc. Surgery T, Aarhus University Hospital, Skejby Sygehus
关键词
blood flow; critical ischaemia; myocardial revascularisation; saphenous vein graft; transit time; ultrasonography;
D O I
10.1016/S1078-5884(96)80282-6
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objectives: To validate the new CardioMed CM 4000 transit time ultrasound apparatus for intraoperative measurement of volume blood flow in vivo in man. Design: Open, prospective series. Materials: Thirteen patients undergoing in situ saphenous vein grafting for chronic critical leg ischaemia and 12 patients subjected to myocardial revascularisation with the internal thoracic artery as coronary bypass. Methods: During operations, volume blood flows were measured simultaneously by exsanguination from the cut distal end of the in situ saphenous vein graft or the internal thoracic artery and by the transit time flowmeter equipment. In addition, the feasibility to detect arteriovenous fistula during in situ saphenous vein grafting was examined. Results: Within the examined blood flow range, the volume blood flow determined by the transit time method corresponded to the directly measured bloodflow. For in situ saphenous vein grafts: y = -2.4 + 0.95 x (r = 0.99; 35 measurements in 13 patients), and for internal thoracic artery grafts: y = -9.6 + 1.1 . x (r = 0.99; 21 measurements in 22 patients), where y is blood flow determined by transit time and x is directly measured blood flow by exsanguination (r = correlation coefficient) as calculated by the least squares regression method. Fistula detection was easy and swift. Conclusions: The transit time apparatus was simple to use during intraoperative settings and gave fast, precise measurements of volume blood flow.
引用
收藏
页码:91 / 96
页数:6
相关论文
共 11 条
[1]  
[Anonymous], 1978, P SAN DIEG BIOM S
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   VALIDATION OF 1-MM AND 2-MM TRANSIT-TIME ULTRASOUND FLOW PROBES ON MESENTERIC-ARTERY AND AORTA OF RATS [J].
DALMEIDA, MS ;
GAUDIN, C ;
LEBREC, D .
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 1995, 268 (03) :H1368-H1372
[4]   PULSED ULTRASONIC TRANSIT TIME FLOWMETER [J].
FRANKLIN, DL ;
BAKER, DW ;
RUSHMER, RF .
IRE TRANSACTIONS ON BIOMEDICAL ELECTRONICS, 1962, BME9 (01) :44-&
[5]  
HARTMAN JC, 1994, J PHARMACOL TOXICOL, V31, P153
[6]   VOLUME BLOOD-FLOW MEASUREMENTS WITH A TRANSIT-TIME FLOWMETER - AN IN-VIVO AND IN-VITRO VARIABILITY AND VALIDATION-STUDY [J].
LUNDELL, A ;
BERGQVIST, D ;
MATTSSON, E ;
NILSSON, B .
CLINICAL PHYSIOLOGY, 1993, 13 (05) :547-557
[7]   COMPARISON OF TRANSIT-TIME AND DOPPLER ULTRASOUND METHODS FOR MEASUREMENT OF FLOW IN AORTOCORONARY BYPASS GRAFTS DURING CARDIAC-SURGERY [J].
MATRE, K ;
BIRKELAND, S ;
HESSEVIK, I ;
SEGADAL, L .
THORACIC AND CARDIOVASCULAR SURGEON, 1994, 42 (03) :170-174
[9]  
RUBERTSSON S, 1993, SURG RES COMM, V13, P309
[10]  
Second European Consensus, 1992, EUR J VASC SURG SA, V6, P1