Ionization and transmission efficiency in an electrospray ionization-mass spectrometry interface

被引:220
作者
Page, Jason S. [1 ]
Kelly, Ryan T. [1 ]
Tang, Keqi [1 ]
Smith, Richard D. [1 ]
机构
[1] Pacific NW Natl Lab, Div Biol Sci, Richland, WA 99352 USA
关键词
D O I
10.1016/j.jasms.2007.05.018
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The ionization and transmission efficiencies of an electrospray ionization (ESI) interface were investigated to advance the understanding of how these factors affect mass spectrometry (MS) sensitivity. In addition, the effects of the ES emitter distance to the inlet, solution flow rate, and inlet temperature were characterized. Quantitative measurements of ES current loss throughout the ESI interface were accomplished by electrically isolating the front surface of the interface from the inner wall of the heated inlet capillary, enabling losses on the two surfaces to be distinguished. In addition, the ES current lost to the front surface of the ESI interface was spatially profiled with a linear array of 340-mu m-diameter electrodes placed adjacent to the inlet capillary entrance. Current transmitted as gas-phase ions was differentiated from charged droplets and solvent clusters by measuring sensitivity with a single quadrupole mass spectrometer. The study revealed a large sampling efficiency into the inlet capillary (> 90% at an emitter distance of I mm), a global rather than a local gas dynamic effect on the shape of the ES plume resulting from the gas flow conductance limit of the inlet capillary, a large (> 80%) loss of analyte ions after transmission through the inlet arising from incomplete desolvation at a solution flow rate of 1.0 mu L/min, and a decrease in analyte ions peak intensity at lower temperatures, despite a large increase in ES current transmission efficiency.
引用
收藏
页码:1582 / 1590
页数:9
相关论文
共 44 条
[1]  
Allanson JP, 1996, RAPID COMMUN MASS SP, V10, P811, DOI 10.1002/(SICI)1097-0231(199605)10:7<811::AID-RCM561>3.0.CO
[2]  
2-Q
[3]   MASS-SPECTROMETRY WITH ION SOURCES OPERATING AT ATMOSPHERIC-PRESSURE [J].
BRUINS, AP .
MASS SPECTROMETRY REVIEWS, 1991, 10 (01) :53-77
[4]   Mechanistic aspects of electrospray ionization [J].
Bruins, AP .
JOURNAL OF CHROMATOGRAPHY A, 1998, 794 (1-2) :345-357
[5]   SPACE-CHARGE-DOMINATED MASS-SPECTROMETRY ION SOURCES - MODELING AND SENSITIVITY [J].
BUSMAN, M ;
SUNNER, J ;
VOGEL, CR .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1991, 2 (01) :1-10
[6]   Practical implications of some recent studies in electrospray ionization fundamentals [J].
Cech, NB ;
Enke, CG .
MASS SPECTROMETRY REVIEWS, 2001, 20 (06) :362-387
[7]   AN ELECTROSPRAY-IONIZATION MASS-SPECTROMETER WITH NEW FEATURES [J].
CHOWDHURY, SK ;
KATTA, V ;
CHAIT, BT .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 1990, 4 (03) :81-87
[8]  
Cole R.B., 1997, Electrospray Ionization Mass Spectrometry
[9]  
Cole RB, 2000, J MASS SPECTROM, V35, P763, DOI 10.1002/1096-9888(200007)35:7<763::AID-JMS16>3.0.CO
[10]  
2-#