Potentiometric biosensors for cholinesterase inhibitor analysis based on mediatorless bioelectrocatalysis

被引:38
作者
Ghindilis, AL [1 ]
Morzunova, TG [1 ]
Barmin, AV [1 ]
Kurochkin, IN [1 ]
机构
[1] RES CTR MOL DIAGNOST & THERAPY,DEPT BIOSENSORS,MOSCOW 113149,RUSSIA
关键词
mediatorless bioelectrocatalysis; butyrylcholinesterase; choline oxidase; peroxidase; inhibitor cholinesterase; biosensor;
D O I
10.1016/0956-5663(96)89436-X
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A potentiometric method for cholinesterase inhibitor analysis based on mediatorless bioelectrocatalysis has been developed. The method includes coimmobilization of three enzymes, butyrylcholinesterase, choline oxidase and peroxidase, on composite carbon electrodes. Catalytic hydrolysis of butyrylcholine and subsequent catalytic oxidation of choline result in the formation of hydrogen peroxide leads to a shift in the electrode potential. The detection limit for trichlorfon analysis is 2 x 10(-13) M. Electrodes remain stable for at least 4 weeks when stored at 277 K. (C) 1996 Elsevier Science Limited.
引用
收藏
页码:873 / 880
页数:8
相关论文
共 21 条
  • [1] DETERMINATION OF BUTYRYLCHOLINESTERASE INHIBITION USING ION TRANSFER ACROSS THE INTERFACE BETWEEN 2 IMMISCIBLE LIQUIDS
    BEATTIE, PD
    INFELTA, PP
    GIRAULT, HH
    [J]. ANALYTICAL CHEMISTRY, 1994, 66 (01) : 52 - 57
  • [2] ANTICHOLINESTERASE ACTIVITY MEASUREMENT BY A CHOLINE BIOSENSOR - APPLICATION IN WATER ANALYSIS
    BERNABEI, M
    CHIAVARINI, S
    CREMISINI, C
    PALLESCHI, G
    [J]. BIOSENSORS & BIOELECTRONICS, 1993, 8 (05) : 265 - 271
  • [3] AN ENZYME AMPEROMETRIC SENSOR FOR TOXICANT DETERMINATION
    BUDNIKOV, HC
    MEDYANTSEVA, EP
    BABKINA, SS
    [J]. JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1991, 310 (1-2): : 49 - 55
  • [4] BUTYRYLCHOLINE ENZYME SENSOR FOR DETERMINING ORGANOPHOSPHORUS INHIBITORS
    CAMPANELLA, L
    ACHILLI, M
    SAMMARTINO, MP
    TOMASSETTI, M
    [J]. BIOELECTROCHEMISTRY AND BIOENERGETICS, 1991, 26 (02): : 237 - 249
  • [5] AMPEROMETRIC SENSORS FOR PEROXIDE, CHOLINE, AND ACETYLCHOLINE BASED ON ELECTRON-TRANSFER BETWEEN HORSERADISH-PEROXIDASE AND A REDOX POLYMER
    GARGUILO, MG
    HUYNH, N
    PROCTOR, A
    MICHAEL, AC
    [J]. ANALYTICAL CHEMISTRY, 1993, 65 (05) : 523 - 528
  • [6] GLUCOSE POTENTIOMETRIC ELECTRODES BASED ON MEDIATORLESS BIOELECTROCATALYSIS - A NEW APPROACH
    GHINDILIS, AL
    KUROCHKIN, IN
    [J]. BIOSENSORS & BIOELECTRONICS, 1994, 9 (4-5) : 353 - 357
  • [7] SENSITIVE DETECTION OF PESTICIDES USING A DIFFERENTIAL ISFET-BASED SYSTEM WITH IMMOBILIZED CHOLINESTERASES
    HENDJI, AMN
    JAFFREZICRENAULT, N
    MARTELET, C
    CLECHET, P
    SHULGA, AA
    STRIKHA, VI
    NETCHIPORUK, LI
    SOLDATKIN, AP
    WLODARSKI, WB
    [J]. ANALYTICA CHIMICA ACTA, 1993, 281 (01) : 3 - 11
  • [8] DETERMINATION OF CHOLINESTERASE IN BLOOD-SERUM WITH A BENZOATE-SENSITIVE MEMBRANE-ELECTRODE
    KATSU, T
    KAYAMOTO, T
    [J]. ANALYTICA CHIMICA ACTA, 1991, 254 (1-2) : 95 - 97
  • [9] BIENZYME SENSORS BASED ON CHEMICALLY MODIFIED ELECTRODES
    KULYS, J
    SCHMID, RD
    [J]. BIOSENSORS & BIOELECTRONICS, 1991, 6 (01) : 43 - 48
  • [10] AMPEROMETRIC ACETYLCHOLINE AND CHOLINE SENSORS WITH IMMOBILIZED ENZYMES
    MASCINI, M
    MOSCONE, D
    [J]. ANALYTICA CHIMICA ACTA, 1986, 179 : 439 - 444