Human factors as a source of error in peak Doppler velocity measurement

被引:121
作者
Lui, EYL
Steinman, AH
Cobbold, RSC
Johnston, KW
机构
[1] Univ Toronto, Inst Biomat & Biomed Engn, Toronto, ON M5S 3G9, Canada
[2] Univ Toronto, Dept Surg, Toronto, ON, Canada
[3] Toronto Gen Hosp, Toronto, ON, Canada
基金
加拿大健康研究院;
关键词
D O I
10.1016/j.jvs.2005.07.014
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objective: The study was conducted to assess the error and variability that results from human factors in Doppler peak velocity measurement. The positioning of the Doppler sample volume in the vessel, adjustment of the Doppler gain and angle, and choice of waveform display size were investigated. We hypothesized that even experienced vascular technologists in a laboratory accredited by the Intersocietal Commission for Accreditation of Vascular Laboratories make significant errors and have significant variability in the subjective adjustments made during measurements. Methods: Problems of patient variability were avoided by having the four technologists measure peak velocities from an in vitro pulsatile flow model with unstenosed and 61% stenosed tubes. To evaluate inaccurate angle and sample volume positioning, a probe holder was used in some of the experiments to fix the Doppler angle at 60 degrees. The effect of Doppler gain was studied at three settings-low, ideal, and saturated gains-that were standardized from the ideal level chosen by consensus amongst the technologists. Two waveform display sizes were also investigated. Peak velocity measurement was assessed by comparison with true peak velocities. For each variable studied, average peak velocities were calculated from the 10 measurements made by each technologist and used to find the percent error from the true value, and the coefficient of variation was used to measure the variability. Results: Doppler angle, sample volume placement, and the Doppler gain were the most significant sources of error and variability. Inaccurate angle and placement increased the variability in measurements from 1% to 2% (range) to 4% to 6% for the straight tube and from 1% to 2% to 3% to 9% for the 61% stenosis. The peak velocity error was increased from 9% to 13% to 7% to 28% for the stenosis. Both measurement error and variability were strongly dependent on the Doppler gain level. At low gain, the error was approximately 10% less than the true value and at saturated gain, 20% greater. The display size only affected measurements from the stenosed tube, increasing the error from 9% to 13% to 15% to 24%. Conclusions: Major factors affecting Doppler peak velocity measurement error and variability were identified. Inaccurate angle and sample volume placement increased the variability. The presence of a stenosis was found to increase the measurement errors. The error was found to depend on the Doppler gain setting, with greater variability at low and saturated gains and on the display size with a stenosis.
引用
收藏
页码:972 / 979
页数:8
相关论文
共 27 条
[1]   Grading carotid stenosis with ultrasound - An interlaboratory comparison [J].
Alexandrov, AV ;
Vital, D ;
Brodie, DS ;
Hamilton, P ;
Grotta, JC .
STROKE, 1997, 28 (06) :1208-1210
[2]  
BAKIRCIOGLU MM, 2004, Patent No. 6733454
[3]   THE EFFECT OF REFRACTION AND ASSUMED SPEEDS OF SOUND IN TISSUE AND BLOOD ON DOPPLER ULTRASOUND BLOOD VELOCITY-MEASUREMENTS [J].
CHRISTOPHER, DA ;
BURNS, PN ;
HUNT, JW ;
FOSTER, FS .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1995, 21 (02) :187-201
[4]  
Chung T. J., 2002, COMPUTATIONAL FLUID
[5]   Interobserver variability of carotid Doppler peak velocity measurcments among technologists in an ICAVL-accredited vascular laboratory [J].
Corriveau, MM ;
Johnston, KW .
JOURNAL OF VASCULAR SURGERY, 2004, 39 (04) :735-741
[6]   Evaluating institutional variability of duplex scanning in the detection of carotid artery stenosis [J].
Criswell, BK ;
Langsfeld, M ;
Tullis, MJ ;
Marek, J .
AMERICAN JOURNAL OF SURGERY, 1998, 176 (06) :591-596
[7]  
Daigle RJ, 1990, J VASC TECHNOL, V14, P206
[9]   Duplex ultrasound criteria for defining the severity of carotid stenosis [J].
Filis, KA ;
Arko, FR ;
Johnson, BL ;
Pipinos, II ;
Harris, EJ ;
Olcott, C ;
Zarins, CK .
ANNALS OF VASCULAR SURGERY, 2002, 16 (04) :413-421
[10]   Ability to use duplex US to quantify internal carotid arterial stenoses: Fact or fiction? [J].
Grant, EG ;
Duerinckx, AJ ;
El Saden, SM ;
Melany, ML ;
Hathout, GM ;
Zimmerman, PT ;
Marumoto, AK ;
Cohen, SN ;
Baker, JD .
RADIOLOGY, 2000, 214 (01) :247-252