A 3D photon superposition/convolution algorithm and its foundation on results of Monte Carlo calculations

被引:184
作者
Ulmer, W
Pyyry, J
Kaissl, W
机构
[1] VARIAN Med Syst Int AG, CH-5405 Baden, Switzerland
[2] Max Planck Inst Phys & Astrophys, Gottingen, Germany
[3] VARIAN Med Syst, Helsinki, Finland
关键词
D O I
10.1088/0031-9155/50/8/010
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Based on previous publications on a triple Gaussian analytical pencil beam model and on Monte Carlo calculations using Monte Carlo codes GEANT-Fluka, versions 95, 98, 2002, and BEAMnrc/EGSnrc, a three-dimensional (3D) superposition/convolution algorithm for photon beams (6 MV, 18 MV) is presented. Tissue heterogeneity is taken into account by electron density information of CT images. A clinical beam consists of a superposition of divergent pencil beams. A slab-geometry was used as a phantom model to test computed results by measurements. An essential result is the existence of further dose build-up and build-down effects in the domain of density discontinuities. These effects have increasing magnitude for field sizes <= 5.5 cm(2) and densities <= 0.25 g cm(-3), in particular with regard to field sizes considered in stereotaxy. They could be confirmed by measurements (mean standard deviation 2%). A practical impact is the dose distribution at transitions from bone to soft tissue, lung or cavities.
引用
收藏
页码:1767 / 1790
页数:24
相关论文
共 40 条
[1]  
AHLSWEDE J, 2000, MED PHYS, V27, P4703
[2]   A PENCIL BEAM MODEL FOR PHOTON DOSE CALCULATION [J].
AHNESJO, A ;
SAXNER, M ;
TREPP, A .
MEDICAL PHYSICS, 1992, 19 (02) :263-273
[3]   Dose calculations for external photon beams in radiotherapy [J].
Ahnesjö, A ;
Aspradakis, MM .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (11) :R99-R155
[4]  
Blomgren H, 1998, Journal of Radiosurgery, V1, P63
[5]   Simulations for inverse radiation therapy treatment planning using a dynamic MLC algorithm [J].
Boman, E ;
Lyyra-Laitinen, T ;
Kolmonen, P ;
Jaatinen, K ;
Tervo, J .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (07) :925-942
[6]   METHODS OF IMAGE-RECONSTRUCTION FROM PROJECTIONS APPLIED TO CONFORMATION RADIOTHERAPY [J].
BORTFELD, T ;
BURKELBACH, J ;
BOESECKE, R ;
SCHLEGEL, W .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (10) :1423-1434
[7]   CALCULATION OF PHOTON DOSE DISTRIBUTIONS IN AN INHOMOGENEOUS-MEDIUM USING CONVOLUTIONS [J].
BOYER, AL ;
MOK, EC .
MEDICAL PHYSICS, 1986, 13 (04) :503-509
[8]   OPTIMIZATION OF STATIONARY AND MOVING BEAM RADIATION-THERAPY TECHNIQUES [J].
BRAHME, A .
RADIOTHERAPY AND ONCOLOGY, 1988, 12 (02) :129-140
[9]   Fast sampling algorithm for the simulation of photon Compton scattering [J].
Brusa, D ;
Stutz, G ;
Riveros, JA ;
FernandezVarea, JM ;
Salvat, F .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 379 (01) :167-175
[10]   Experimental determination of the dose kernel in high-energy x-ray beams [J].
Ceberg, CP ;
Bjarngard, BE ;