Simulations for inverse radiation therapy treatment planning using a dynamic MLC algorithm

被引:3
作者
Boman, E [1 ]
Lyyra-Laitinen, T
Kolmonen, P
Jaatinen, K
Tervo, J
机构
[1] Univ Kuopio, Dept Appl Phys, Res Inst Radiotherapy Phys, FIN-70211 Kuopio, Finland
[2] Kuopio Univ Hosp, Dept Clin Physiol & Nucl Med, Kuopio, Finland
[3] Univ Kuopio, Dept Comp Sci & Appl Math, Kuopio, Finland
关键词
D O I
10.1088/0031-9155/48/7/309
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The inverse radiation treatment planning model for a dynamic multileaf collimator (MLC) is used to find the optimal solution of planning problem. The model for dynamic MLC is explained in Tervo et al (2003 Appl. Math. Comput. 135 227-50). The advantage of this model is that it optimizes leaf velocity parameters directly. Our algorithm uses a gradient-based local optimization method. Two patient cases, prostate carcinoma and tonsilla carcinoma, are studied. Field arrangements are pre-selected and velocity parameters for MLC leaves are optimized to obtain the prescribed dose in the patient space. In both simulated cases, high dose distribution conforms the planning target volume well and organs-at-risk are saved in most parts. Simulations show that the model has its functionality in patient treatments, although it is still formal and needs further development.
引用
收藏
页码:925 / 942
页数:18
相关论文
共 41 条
[1]  
[Anonymous], 1995, HDB GLOBAL OPTIMIZAT
[2]  
BORGERS C, 1997, COMPUTATIONAL RADIOL
[3]   METHODS OF IMAGE-RECONSTRUCTION FROM PROJECTIONS APPLIED TO CONFORMATION RADIOTHERAPY [J].
BORTFELD, T ;
BURKELBACH, J ;
BOESECKE, R ;
SCHLEGEL, W .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (10) :1423-1434
[4]   X-RAY FIELD COMPENSATION WITH MULTILEAF COLLIMATORS [J].
BORTFELD, TR ;
KAHLER, DL ;
WALDRON, TJ ;
BOYER, AL .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 28 (03) :723-730
[5]   Intensity-modulated radiation therapy with dynamic multileaf collimators [J].
Boyer, AL ;
Yu, CX .
SEMINARS IN RADIATION ONCOLOGY, 1999, 9 (01) :48-59
[6]  
BRAHME A, 1995, RAD THERAPY PHYSICS, P209
[7]   ON THE USE OF CIMMINO SIMULTANEOUS PROJECTIONS METHOD FOR COMPUTING A SOLUTION OF THE INVERSE PROBLEM IN RADIATION-THERAPY TREATMENT PLANNING [J].
CENSOR, Y ;
ALTSCHULER, MD ;
POWLIS, WD .
INVERSE PROBLEMS, 1988, 4 (03) :607-623
[8]  
Censor Y, 1997, PARALLEL OPTIMIZATIO
[9]   Optimization of intensity modulated beams with volume constraints using two methods: Cost function minimization and projections onto convex sets [J].
Cho, PS ;
Lee, S ;
Marks, RJ ;
Oh, SH ;
Sutlief, SG ;
Phillips, MH .
MEDICAL PHYSICS, 1998, 25 (04) :435-443
[10]   Hardware-sensitive optimization for intensity modulated radiotherapy [J].
Cho, PS ;
Marks, RJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (02) :429-440