Accuracy and repeatability of pediatric cardiac output measurement using Doppler: 20-year review of the literature

被引:110
作者
Chew, MS [1 ]
Poelaert, J
机构
[1] Univ Lund Hosp, Dept Anesthesia & Intens Care, S-22185 Lund, Sweden
[2] State Univ Ghent Hosp, Dept Intens Care, B-9000 Ghent, Belgium
关键词
pediatric; Doppler; cardiac output; accuracy; reproducibility;
D O I
10.1007/s00134-003-1967-9
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective. Review of the accuracy and repeatability of Doppler cardiac output (CO) measurements in children. <LF>Design. Publications in the scientific literature retrieved using a computerized Medline search from 1982-2002 and a manual review of article bibliographies. Studies comparing Doppler flow measurements with thermodilution, Fick, or dye dilution methods in the pediatric critical care setting were identified to assess the bias, precision, and intra- and interobserver repeatability of Doppler CO measurement. Where results were not suitable for comparison and the original measurements available, data were re-analyzed using appropriate statistical methods and presented in comparative tables. Results. The precision of pediatric Doppler CO measurements compared to thermodilution, dye dilution, or Fick methods is around 30% and repeatability varies from less than 1% to 22%. Bias is generally less than 10% but varies considerably. Conclusions. The bias, precision, and repeatability from study to study indicate that Doppler CO measurements are acceptably reproducible in children, with best results when used to track changes rather than absolute values, and using the transesophageal approach.
引用
收藏
页码:1889 / 1894
页数:6
相关论文
共 32 条
[1]   NON-INVASIVE PULSED DOPPLER DETERMINATION OF CARDIAC-OUTPUT IN NEONATES AND CHILDREN [J].
ALVERSON, DC ;
ELDRIDGE, M ;
DILLON, T ;
YABEK, SM ;
BERMAN, W .
JOURNAL OF PEDIATRICS, 1982, 101 (01) :46-50
[2]   PROSPECTIVE, RANDOMIZED TRIAL OF SURVIVOR VALUES OF CARDIAC INDEX, OXYGEN DELIVERY, AND OXYGEN-CONSUMPTION AS RESUSCITATION END-POINTS IN SEVERE TRAUMA [J].
BISHOP, MH ;
SHOEMAKER, WC ;
APPEL, PL ;
MEADE, P ;
ORDOG, GJ ;
WASSERBERGER, J ;
WO, CJ ;
RIMLE, DA ;
KRAM, HB ;
UMALI, R ;
KENNEDY, F ;
SHULESHKO, J ;
STEPHEN, CM ;
SHORI, SK ;
THADEPALLI, HD .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1995, 38 (05) :780-787
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   A RANDOMIZED CLINICAL-TRIAL OF THE EFFECT OF DELIBERATE PERIOPERATIVE INCREASE OF OXYGEN DELIVERY ON MORTALITY IN HIGH-RISK SURGICAL PATIENTS [J].
BOYD, O ;
GROUNDS, RM ;
BENNETT, ED .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1993, 270 (22) :2699-2707
[5]   Pediatric cardiac output measurement using surface integration of velocity vectors:: An in vivo validation study [J].
Chew, MS ;
Brandberg, J ;
Bjarum, S ;
Baek-Jensen, K ;
Sloth, E ;
Hasenkam, JM ;
Janerot-Sjöberg, B .
CRITICAL CARE MEDICINE, 2000, 28 (11) :3664-3671
[6]   Suprasternal Doppler ultrasound for assessment of stroke distance [J].
Childs, C ;
Goldring, S ;
Tann, W ;
Hillier, VF .
ARCHIVES OF DISEASE IN CHILDHOOD, 1998, 79 (03) :251-255
[7]   CARDIAC-OUTPUT DETERMINATIONS IN THE NEWBORN - REPRODUCIBILITY OF THE PULSED DOPPLER VELOCITY-MEASUREMENT [J].
CLAFLIN, KS ;
ALVERSON, DC ;
PATHAK, D ;
ANGELUS, P ;
BACKSTROM, C ;
WERNER, S .
JOURNAL OF ULTRASOUND IN MEDICINE, 1988, 7 (06) :311-315
[8]   A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques [J].
Critchley, LAH ;
Critchley, JAJH .
JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1999, 15 (02) :85-91
[9]  
DARMON PL, 1994, ANESTHESIOLOGY, V80, P796, DOI 10.1097/00000542-199404000-00011
[10]   A TRIAL OF GOAL-ORIENTED HEMODYNAMIC THERAPY IN CRITICALLY ILL PATIENTS [J].
GATTINONI, L ;
BRAZZI, L ;
PELOSI, P ;
LATINI, R ;
TOGNONI, G ;
PESENTI, A ;
FUMAGALLI, R .
NEW ENGLAND JOURNAL OF MEDICINE, 1995, 333 (16) :1025-1032