Intravascular ultrasound assessment of lesions with target vessel failure after sirolimus-eluting stent implantation

被引:75
作者
Takebayashi, H [1 ]
Kobayashi, Y [1 ]
Mintz, GS [1 ]
Carlier, SG [1 ]
Fujii, K [1 ]
Yasuda, T [1 ]
Moussa, I [1 ]
Mehran, R [1 ]
Dangas, GD [1 ]
Collins, MB [1 ]
Kreps, E [1 ]
Lansky, AJ [1 ]
Stone, GW [1 ]
Leon, MB [1 ]
Moses, JW [1 ]
机构
[1] Cardiovasc Res Fdn, New York, NY USA
关键词
D O I
10.1016/j.amjcard.2004.10.020
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Intravascular ultrasound (IVUS) evaluation was performed in 33 lesions with sirolimus-eluting stent (SES) failure: 4 thromboses; 26 in-stent restenoses (including 6 edge stenoses), 4 new stenoses >5 mm proximal to the stent, and 1 patient with no evidence of the implanted SES (presumably because of embolization): A minimum stent area <5.0 mm(2) (stent underexpansion) was observed in 67% of all SES failures (in particular, 67% of intrastent restenosis); negative remodeling was observed in 4 of 6 stent edge restenoses, and new lesions were secondary to an increase in plaque area. (C) 2005 by Excerpta Medica Inc.
引用
收藏
页码:498 / 502
页数:5
相关论文
共 14 条
[1]   The contribution of "mechanical" problems to in-stent restenosis: An intravascular ultrasonographic analysis of 1090 consecutive in-stent restenosis lesions [J].
Castagna, MT ;
Mintz, GS ;
Leiboff, BO ;
Ahmed, JM ;
Mehran, R ;
Satler, LF ;
Kent, KM ;
Pichard, AD ;
Weissman, NJ .
AMERICAN HEART JOURNAL, 2001, 142 (06) :970-974
[2]   Predictors of subacute stent thrombosis - Results of a systematic intravascular ultrasound study [J].
Cheneau, E ;
Leborgne, L ;
Mintz, GS ;
Kotani, J ;
Pichard, AD ;
Satler, LF ;
Canos, D ;
Castagna, M ;
Weissman, NJ ;
Waksman, R .
CIRCULATION, 2003, 108 (01) :43-47
[3]   Randomized study to assess the effectiveness of slow- and moderate-release polymer-based paclitaxel-eluting stents for coronary artery lesions [J].
Colombo, A ;
Drzewiecki, J ;
Banning, A ;
Grube, E ;
Hauptmann, K ;
Silber, S ;
Dudek, D ;
Fort, S ;
Schiele, F ;
Zmudka, K ;
Guagliumi, G ;
Russell, ME .
CIRCULATION, 2003, 108 (07) :788-794
[4]   Randomized study to evaluate sirolimus-eluting stents implanted at coronary bifurcation lesions [J].
Colombo, A ;
Moses, JW ;
Morice, MC ;
Ludwig, J ;
Holmes, DR ;
Spanos, V ;
Louvard, Y ;
Desmedt, B ;
Di Mario, C ;
Leon, MB .
CIRCULATION, 2004, 109 (10) :1244-1249
[5]   Patterns and mechanisms of in-stent restenosis - A serial intravascular ultrasound study [J].
Hoffmann, R ;
Mintz, GS ;
Dussaillant, GR ;
Popma, JJ ;
Pichard, AD ;
Satler, LF ;
Kent, KM ;
Griffin, J ;
Leon, MB .
CIRCULATION, 1996, 94 (06) :1247-1254
[6]   Coronary restenosis after sirolimus-eluting stent implantation - Morphological description and mechanistic analysis from a consecutive series of cases [J].
Lemos, PA ;
Saia, F ;
Ligthart, JMR ;
Arampatzis, CA ;
Sianos, G ;
Tanabe, K ;
Hoye, A ;
Degertekin, M ;
Daemen, J ;
McFadden, E ;
Hofma, S ;
Smits, PC ;
de Feyter, P ;
van der Giessen, WJ ;
van Domburg, RT ;
Serruys, PW .
CIRCULATION, 2003, 108 (03) :257-260
[7]   Angiographic patterns of in-stent restenosis - Classification and implications for long-term outcome [J].
Mehran, R ;
Dangas, G ;
Abizaid, AS ;
Mintz, GS ;
Lansky, AJ ;
Satler, LF ;
Pichard, AD ;
Kent, KM ;
Stone, GW ;
Leon, MB .
CIRCULATION, 1999, 100 (18) :1872-1878
[8]   American College of Cardiology Clinical Expert Consensus Document on Standards for Acquisition, Measurement and Reporting of Intravascular Ultrasound Studies (IVUS) - A report of the American College of Cardiology Task Force on Clinical Expert Consensus - Documents developed in collaboration with the European Society of Cardiology endorsed by the Society of Cardiac Angiography and Interventions [J].
Mintz, GS ;
Nissen, SE ;
Anderson, WD ;
Bailey, SR ;
Erbel, R ;
Fitzgerald, PJ ;
Pinto, FJ ;
Rosenfield, K ;
Siegel, RJ ;
Tuzcu, EM ;
Yock, PG .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 37 (05) :1478-1492
[9]   A randomized comparison of a sirolimus-eluting stent with a standard stent for coronary revascularization. [J].
Morice, M ;
Serruys, PW ;
Sousa, JE ;
Fajadet, J ;
Hayashi, EB ;
Perin, M ;
Colombo, A ;
Schuler, G ;
Barragan, P ;
Guagliumi, G ;
Molnar, F ;
Falotico, R .
NEW ENGLAND JOURNAL OF MEDICINE, 2002, 346 (23) :1773-1780
[10]   Sirolimus-eluting stents versus standard stents in patients with stenosis in a native coronary artery [J].
Moses, JW ;
Leon, MB ;
Popma, JJ ;
Fitzgerald, PJ ;
Holmes, DR ;
O'Shaughnessy, C ;
Caputo, RP ;
Kereiakes, DJ ;
Williams, DO ;
Teirstein, PS ;
Jaeger, JL ;
Kuntz, RE .
NEW ENGLAND JOURNAL OF MEDICINE, 2003, 349 (14) :1315-1323