Patterns and mechanisms of in-stent restenosis - A serial intravascular ultrasound study

被引:1036
作者
Hoffmann, R
Mintz, GS
Dussaillant, GR
Popma, JJ
Pichard, AD
Satler, LF
Kent, KM
Griffin, J
Leon, MB
机构
[1] WASHINGTON HOSP CTR,INTRAVASCULAR ULTRASOUND IMAGING LAB,WASHINGTON,DC 20010
[2] WASHINGTON HOSP CTR,CARDIAC CATHETERIZAT LAB,WASHINGTON,DC 20010
关键词
stents; ultrasonics; restenosis; angioplasty;
D O I
10.1161/01.CIR.94.6.1247
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Studies have suggested that restenosis within Palmaz-Schatz stents results from neointimal hyperplasia or chronic stent recoil and occurs more frequently at the articulation. Methods and Results Serial intravascular ultrasound (IVUS) was performed after intervention and at follow-up in 142 stents in 115 lesions. IVUS measurements (external elastic membrane [EEM], stent, and lumen cross-sectional areas [CSAs] and diameters) were performed, and plaque CSA (EEM lumen in reference segments and stent lumen in stented segments), late lumen loss (Delta lumen), remodeling (Delta EEM in reference segments and Delta stent in stented segments), and tissue growth (Delta plaque) were calculated. After intervention, the lumen tended to be smallest at the articulation because of tissue prolapse. At follow-up, tissue growth was uniformly distributed throughout the stent; the tendency for greater neointimal tissue accumulation at the central articulation reached statistical significance only when normalized for the smaller postintervention lumen CSA. Ln stented segments, late lumen area loss correlated strongly with tissue growth but only weakly with remodeling. Stents affected adjacent vessel segments; remodeling progressively increased and tissue growth progressively decreased at distances from the edge of the stent. These findings were similar in native arteries and saphenous vein grafts and in lesions treated with one or two stents. There was no difference in the postintervention or follow-up lumen (at the junction of the two stents) when overlapped were compared with nonoverlapped stents. Conclusions Late lumen loss and in-stent restenosis were the result of neointimal tissue proliferation, which tended to be uniformly distributed over the length of the stent.
引用
收藏
页码:1247 / 1254
页数:8
相关论文
共 29 条
  • [1] Bailey Steven R., 1994, P893
  • [2] INTRACORONARY STENTING WITHOUT ANTICOAGULATION ACCOMPLISHED WITH INTRAVASCULAR ULTRASOUND GUIDANCE
    COLOMBO, A
    HALL, P
    NAKAMURA, S
    ALMAGOR, Y
    MAIELLO, L
    MARTINI, G
    GAGLIONE, A
    GOLDBERG, SL
    TOBIS, JM
    [J]. CIRCULATION, 1995, 91 (06) : 1676 - 1688
  • [3] SMALL STENT SIZE AND INTIMAL HYPERPLASIA CONTRIBUTE TO RESTENOSIS - A VOLUMETRIC INTRAVASCULAR ULTRASOUND ANALYSIS
    DUSSAILLANT, GR
    MINTZ, GS
    PICHARD, AD
    KENT, KM
    SATLER, LF
    POPMA, JJ
    WONG, SC
    LEON, MB
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1995, 26 (03) : 720 - 724
  • [4] A RANDOMIZED COMPARISON OF CORONARY-STENT PLACEMENT AND BALLOON ANGIOPLASTY IN THE TREATMENT OF CORONARY-ARTERY DISEASE
    FISCHMAN, DL
    LEON, MB
    BAIM, DS
    SCHATZ, RA
    SAVAGE, MP
    PENN, I
    DETRE, K
    VELTRI, L
    RICCI, D
    NOBUYOSHI, M
    CLEMAN, M
    HEUSER, R
    ALMOND, D
    TEIRSTEIN, PS
    FISH, RD
    COLOMBO, A
    BRINKER, J
    MOSES, J
    SHAKNOVICH, A
    HIRSHFELD, J
    BAILEY, S
    ELLIS, S
    RAKE, R
    GOLDBERG, S
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 1994, 331 (08) : 496 - 501
  • [5] BENEFIT OF INTRACORONARY ULTRASOUND IN THE DEPLOYMENT OF PALMAZ-SCHATZ STENTS
    GOLDBERG, SL
    COLOMBO, A
    NAKAMURA, S
    ALMAGOR, Y
    MAIELLO, L
    TOBIS, JM
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 24 (04) : 996 - 1003
  • [6] MECHANISMS OF RESTENOSIS AND REDILATION WITHIN CORONARY STENTS - QUANTITATIVE ANGIOGRAPHIC ASSESSMENT
    GORDON, PC
    GIBSON, CM
    COHEN, DJ
    CARROZZA, JP
    KUNTZ, RE
    BAIM, DS
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 21 (05) : 1166 - 1174
  • [7] INTRAVASCULAR ULTRASOUND AFTER LOW AND HIGH INFLATION PRESSURE CORONARY-ARTERY STENT IMPLANTATION
    GORGE, G
    HAUDE, M
    VOEGELE, E
    GERBER, T
    RUPPRECHT, HJ
    MEYER, J
    ERBEL, R
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1995, 26 (03) : 725 - 730
  • [8] HAUDE M, 1991, BRIT HEART J, V66, P337
  • [9] HODGSON JM, 1989, INT J CARD IMAGING, V4, P186
  • [10] LUMINAL LOSS AND SITE OF RESTENOSIS AFTER PALMAZ-SCHATZ CORONARY STENT IMPLANTATION
    IKARI, Y
    HARA, K
    TAMURA, T
    SAEKI, F
    YAMAGUCHI, T
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1995, 76 (03) : 117 - 120