Development of a coupled-column liquid chromatographic tandem mass spectrometric method for the direct determination of betamethasone in urine

被引:40
作者
Polettini, A [1 ]
Bouland, GM [1 ]
Montagna, M [1 ]
机构
[1] Univ Pavia, Dept Legal Med & Publ Hlth, I-27100 Pavia, Italy
来源
JOURNAL OF CHROMATOGRAPHY B | 1998年 / 713卷 / 02期
关键词
mass spectrometry; betamethasone;
D O I
10.1016/S0378-4347(98)00177-7
中图分类号
Q5 [生物化学];
学科分类号
071010 [生物化学与分子生物学]; 081704 [应用化学];
摘要
Different hyphenated liquid chromatographic (LC) and mass spectrometric (MS) techniques were investigated in,order to set-up a method for the fast, direct analysis of betamethasone in hydrolysed and non-hydrolysed urine using large-volume sample injection. After the optimisation of the LC parameters using a traditional UV detector and of the thermospray and mass spectrometric parameters by flow injection, urine samples (0.5 mi) were submitted to analysis by either LC combined with tandem mass spectrometry (MS-MS), coupled-column LC (LC-LC) combined with single quadrupole MS, and LC-LC-MS-MS. Both the three-step configurations (LC-MS-MS and LC-LC-MS) did not provide satisfactory results: loss of sensitivity was noted in the case of LC-MS-MS (likely due to reduced efficiency in the ionisation of betamethasone in the thermospray owing to the presence of large amounts of matrix interference), while in the case of LC-LC-MS a high chemical noise resulting in insufficient selectivity of detection was observed. On the contrary, LC-LC-MS-MS analysis proved to meet the demand of high speed of analysis (sample throughput, 4.5 h(-1)), selectivity, and sensitivity (LOQ, 1 ng/ml; LOD, 0.2 ng/ml). Notwithstanding the complex analytical system adopted, the developed procedure was manageable and very robust, provided that at the beginning of each analytical session the performance of the system was controlled by checking the retention time of the analytes on the first analytical column with UV detection and by optimising vaporiser temperature of the thermospray by flow injection. (C) 1998 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:339 / 352
页数:14
相关论文
共 21 条
[1]
COMBINED LIQUID-CHROMATOGRAPHY MASS-SPECTROMETRY .2. TECHNIQUES AND MECHANISMS OF THERMOSPRAY [J].
ARPINO, P .
MASS SPECTROMETRY REVIEWS, 1990, 9 (06) :631-669
[2]
Analysis of dexamethasone and betamethasone in bovine urine by purification with an ''on-line'' immunoaffinity chromatography high-performance liquid chromatography system and determination by gas chromatography mass spectrometry [J].
Bagnati, R ;
Ramazza, V ;
Zucchi, M ;
Simonella, A ;
Leone, F ;
Bellini, A ;
Fanelli, R .
ANALYTICAL BIOCHEMISTRY, 1996, 235 (02) :119-126
[3]
*COUNC EUR COMM, 1996, 86469EEC
[4]
DETERMINATION OF DEXAMETHASONE IN URINE AND FECES OF TREATED CATTLE WITH NEGATIVE CHEMICAL-IONIZATION - MASS-SPECTROMETRY [J].
COURTHEYN, D ;
VERCAMMEN, J ;
DEBRABANDER, H ;
VANDENREYT, I ;
BATJOENS, P ;
VANOOSTHUYZE, K ;
VANPETEGHEM, C .
ANALYST, 1994, 119 (12) :2557-2564
[5]
DETERMINATION OF DRUG ENANTIOMERS IN BIOLOGICAL SAMPLES BY COUPLED COLUMN LIQUID-CHROMATOGRAPHY AND LIQUID-CHROMATOGRAPHY MASS-SPECTROMETRY [J].
EDHOLM, LE ;
LINDBERG, C ;
PAULSON, J ;
WALHAGEN, A .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1988, 424 (01) :61-72
[6]
RAPID-DETERMINATION OF METHANDROSTENOLONE IN EQUINE URINE BY ISOTOPE-DILUTION LIQUID-CHROMATOGRAPHY TANDEM MASS-SPECTROMETRY [J].
EDLUND, PO ;
BOWERS, L ;
HENION, J ;
COVEY, TR .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1989, 497 :49-57
[7]
DETERMINATION OF SERUM CORTISOL BY THERMOSPRAY LIQUID-CHROMATOGRAPHY MASS-SPECTROMETRY - COMPARISON WITH GAS-CHROMATOGRAPHY MASS-SPECTROMETRY [J].
GASKELL, SJ ;
ROLLINS, K ;
SMITH, RW ;
PARKER, CE .
BIOMEDICAL AND ENVIRONMENTAL MASS SPECTROMETRY, 1987, 14 (12) :717-722
[8]
SIMULTANEOUS DETERMINATION OF BECLOMETHASONE, BECLOMETHASONE MONOPROPIONATE AND BECLOMETHASONE DIPROPIONATE IN BIOLOGICAL-FLUIDS USING A PARTICLE BEAM INTERFACE FOR COMBINING LIQUID-CHROMATOGRAPHY WITH NEGATIVE-ION CHEMICAL IONIZATION MASS-SPECTROMETRY [J].
GIRAULT, J ;
ISTIN, B ;
MALGOUYAT, JM ;
BRISSON, AM ;
FOURTILLAN, JB .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1991, 564 (01) :43-53
[9]
A RAPID AND HIGHLY SENSITIVE METHOD FOR THE QUANTITATIVE-DETERMINATION OF DEXAMETHASONE IN PLASMA, SYNOVIAL-FLUID AND TISSUES BY COMBINED GAS-CHROMATOGRAPHY NEGATIVE-ION CHEMICAL IONIZATION MASS-SPECTROMETRY [J].
GIRAULT, J ;
ISTIN, B ;
FOURTILLAN, JB .
BIOMEDICAL AND ENVIRONMENTAL MASS SPECTROMETRY, 1990, 19 (05) :295-302
[10]
Haynes R.C. Jr, 1992, GOODMAN GILMANNS PHA, P1431