Dynamic liquid-liquid-liquid microextraction with automated movement of the acceptor phase

被引:63
作者
Jiang, XM [1 ]
Oh, SY [1 ]
Lee, HK [1 ]
机构
[1] Natl Univ Singapore, Dept Chem, Singapore 117543, Singapore
关键词
D O I
10.1021/ac040153v
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new dynamic liquid-liquid-liquid microextraction procedure, with the automated movement of acceptor phase (LLLME/AMAP) to facilitate mass transfer, was developed in this study. Four compounds, 3-nitrophenol, 4-nitrophenol, 3,4-dinitrophenol, and 2,4-dichlorophenol, were used as model compounds to be preconcentrated from water samples. The extraction involved filling a 2-cm length of hollow fiber with 4 mu L of acceptor solution using a conventional microsyringe, followed by impregnation of the pores of the fiber wall with 1-octanol. The fiber was then immersed in 4 mL of aqueous sample solution. The analytes in the sample solution were extracted into the organic solvent and then back-extracted into the acceptor solution. During extraction, the acceptor phase was repeatedly moved in and out of the hollow fiber channel and the syringe controlled by a syringe pump. Separation and quantitative analyses were then performed by using high-performance liquid chromatography. The results indicated that up to 400-fold enrichment of the analytes could be obtained under the optimized conditions. The enrichment factors were two times those of static liquid-liquid-liquid microextraction. Good repeatabilities (RSD values below 9.30%) were obtained. The calibration linear range was from 10 to 1000 ng/mL with the square of the correlation coefficient (r(2)) > 0.9916. Detection limits were in the range of 0.45-0.98 ng/mL. In addition, as compared with the previously reported dynamic three-phase microextraction in which there was no relative movement between the acceptor and the organic phase (which is not conducive to effective mass transfer), this new method shows much higher extraction efficiency. All these results suggest that this new dynamic LLLME/AMAP technique could be a better alternative to the previous LLLME for the extraction of analytes from aqueous samples.
引用
收藏
页码:1689 / 1695
页数:7
相关论文
共 19 条
[1]  
CUSSLER EL, 1997, DIFFUSION MASS TRANS, pCH13
[2]  
CUSSLER EL, 1997, DIFFUSION MASS TRANS, pCH16
[3]   CONFIRMATION OF CHLOROTRIAZINE PESTICIDES, THEIR DEGRADATION PRODUCTS AND ORGANOPHOSPHORUS PESTICIDES IN SOIL SAMPLES USING GAS-CHROMATOGRAPHY MASS-SPECTROMETRY WITH ELECTRON-IMPACT AND POSITIVE-ION AND NEGATIVE-ION CHEMICAL IONIZATION [J].
DURAND, G ;
BARCELO, D .
ANALYTICA CHIMICA ACTA, 1991, 243 (02) :259-271
[4]   DETERMINATION OF ORGANIC POLLUTANTS IN SMALL SAMPLES OF GROUNDWATERS BY LIQUID-LIQUID-EXTRACTION AND CAPILLARY GAS-CHROMATOGRAPHY [J].
HARRISON, I ;
LEADER, RU ;
HIGGO, JJW ;
TJELL, JC .
JOURNAL OF CHROMATOGRAPHY A, 1994, 688 (1-2) :181-188
[5]   Liquid phase microextraction in a single drop of organic solvent by using a conventional microsyringe [J].
He, Y ;
Lee, HK .
ANALYTICAL CHEMISTRY, 1997, 69 (22) :4634-4640
[6]   Dynamic three-phase microextraction as a sample preparation technique prior to capillary electrophoresis [J].
Hou, L ;
Lee, HK .
ANALYTICAL CHEMISTRY, 2003, 75 (11) :2784-2789
[7]   Solvent microextraction into a single drop [J].
Jeannot, MA ;
Cantwell, FF .
ANALYTICAL CHEMISTRY, 1996, 68 (13) :2236-2240
[8]   Mass transfer characteristics of solvent extraction into a single drop at the tip of a syringe needle [J].
Jeannot, MA ;
Cantwell, FF .
ANALYTICAL CHEMISTRY, 1997, 69 (02) :235-239
[9]   LIQUID-LIQUID-EXTRACTION AND ANALYSIS OF PARACETAMOL (ACETAMINOPHEN) AND ITS MAJOR METABOLITES IN BIOLOGICAL-FLUIDS BY REVERSED-PHASE ION-PAIR CHROMATOGRAPHY [J].
KAMALI, F ;
HERD, B .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1990, 530 (01) :222-225
[10]   Solvent microextraction with simultaneous back-extraction for sample cleanup and preconcentration: Preconcentration into a single microdrop [J].
Ma, MH ;
Cantwell, FF .
ANALYTICAL CHEMISTRY, 1999, 71 (02) :388-393