Analysis of eighteen antidepressants, four atypical antipsychotics and active metabolites in serum by liquid chromatography:: a simple tool for therapeutic drug monitoring

被引:265
作者
Frahnert, C [1 ]
Rao, ML [1 ]
Grasmäder, K [1 ]
机构
[1] Univ Bonn, Dept Psychiat, D-53105 Bonn, Germany
来源
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES | 2003年 / 794卷 / 01期
关键词
therapeutic drug monitoring; antidepressants; antipsychotics; atypical;
D O I
10.1016/S1570-0232(03)00393-3
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Therapeutic drug monitoring necessitates efficient, fast and reliable analytical methods validated by external quality control. We therefore devised an isocratic reversed-phase HPLC method with ultraviolet detection and optimised this to quantify mirtazapine, reboxetine, moclobemide, venlafaxine, O-desmethylvenlafaxine, paroxetine, fluvoxamine, fluoxetine, norfluoxetine. sertraline. citalopram, armtriptyline, nortriptyline, imipramine, desipramine, doxepin, nordoxepin, clomipramine, norclomipramine, trimipramine, mianserine, maprotiline, normaprotiline, amisulpride, clozapine, norclozapine, quetiapine. risperidone and 9-OH-risperidone in human serum. After solid-phase extraction of the drugs and metabolites, the chromatographic separation was achieved on a Nucleosil 100-Protect 1 column with acetonitrile-potassium dihydrogenphosphate buffer as mobile phase. The method was validated for therapeutic and toxic serum ranges. A linear relationship (r>0.998) was obtained between the concentration and the detector signal. Recoveries were between 75 and 99% for the drugs and metabolites. The accuracy of the quality control samples, expressed as percent recovery, ranged from 91 to 118%; intra- and inter-assay-relative standard deviations were 0.9-10.2% and 0.9-9.7%, respectively. Additional external quality control is carried out since 3 years. This method is applicable to rapidly and effectively analyze serum or plasma samples for therapeutic drug monitoring of about 30 antidepressants and atypical antipsychotics. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:35 / 47
页数:13
相关论文
共 39 条
[21]   THE RELATIONSHIP BETWEEN ANTIDEPRESSANT RESPONSE AND TRICYCLIC ANTIDEPRESSANT PLASMA-CONCENTRATIONS - A RETROSPECTIVE ANALYSIS OF THE LITERATURE USING LOGISTIC-REGRESSION ANALYSIS [J].
PERRY, PJ ;
PFOHL, BM ;
HOLSTAD, SG .
CLINICAL PHARMACOKINETICS, 1987, 13 (06) :381-392
[22]  
PRESKORN SH, 1991, J CLIN PSYCHIAT, V52, P23
[23]  
Queiroz R H, 1995, Pharm Acta Helv, V70, P181, DOI 10.1016/0031-6865(95)00019-6
[24]  
RAO ML, 1986, PSYCHOPHARMACOLOGY, V90, P548
[25]  
Rao ML, 2001, FORTSCHR NEUROL PSYC, V69, P510
[26]   Is therapeutic drug monitoring a case for optimizing clinical outcome and avoiding interactions of the selective serotonin reuptake inhibitors? [J].
Rasmussen, BB ;
Brosen, K .
THERAPEUTIC DRUG MONITORING, 2000, 22 (02) :143-154
[27]   Automated determination of amisulpride by liquid chromatography with column switching and spectrophotometric detection [J].
Sachse, J ;
Härtter, S ;
Weigmann, H ;
Hiemke, C .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2003, 784 (02) :405-410
[28]   RAPID AND SIMPLE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF TRICYCLIC ANTIDEPRESSANTS FOR ROUTINE AND EMERGENCY SERUM ANALYSIS [J].
SEGATTI, MP ;
NISI, G ;
GROSSI, F ;
MANGIAROTTI, M ;
LUCARELLI, C .
JOURNAL OF CHROMATOGRAPHY, 1991, 536 (1-2) :319-325
[29]   Relationship between plasma concentrations of clozapine and norclozapine and therapeutic response in patients with schizophrenia resistant to conventional neuroleptics [J].
Spina, E ;
Avenoso, A ;
Facciolà, G ;
Scordo, MG ;
Ancione, M ;
Madia, AG ;
Ventimiglia, A ;
Perucca, E .
PSYCHOPHARMACOLOGY, 2000, 148 (01) :83-89
[30]   Plasma concentrations of risperidone and 9-hydroxyrisperidone during combined treatment with paroxetine [J].
Spina, E ;
Avenoso, A ;
Facciolà, G ;
Scordo, MG ;
Ancione, M ;
Madia, A .
THERAPEUTIC DRUG MONITORING, 2001, 23 (03) :223-227