Discrimination between normal and early glaucomatous eyes with scanning laser polarimeter with fixed and variable corneal compensator settings

被引:24
作者
Brusini, P [1 ]
Salvetat, ML [1 ]
Parisi, L [1 ]
Zeppieri, M [1 ]
Tosoni, C [1 ]
机构
[1] Santa Maria Misericordia Hosp, Dept Ophthalmol, I-33100 Udine, Italy
关键词
early primary open-angle glaucoma; scanning laser polarimetry; fixed corneal compensator; variable corneal compensator; sensitivity; specificity;
D O I
10.1177/112067210501500409
中图分类号
R77 [眼科学];
学科分类号
100212 [眼科学];
摘要
PURPOSE. To evaluate the ability of scanning laser polarimetry (SLP) with a fixed corneal polarization compensator (GDx-FCC Nerve Fiber Analyzer) compared to one with a variable one (GDx-VCC) in the discrimination between healthy and early glaucomatous eyes. METHODS. Forty patients with early glaucomatous visual field defects, having a mean deviation of 3.1 +/- 1.6 dB and a pattern standard deviation of 3.1 +/- 0.9 dB, and 40 controls underwent both GDx-FCC and GDx-VCC. One eye per patient was considered. The cut-off point, taken as the value dividing healthy from glaucomatous eyes with highest probability, was determined for each GDx parameter. Linear discriminant functions (LDFs) were separately developed for GDx-FCC and GDx-VCC parameters. Sensitivity, specificity, and area under the receiver operating characteristic curve (AROC) for discriminating between healthy and glaucomatous eyes were calculated for each GDx parameter, both according to the GDx normative database and after the selection of new cut-off points, and for the LDFs. RESULTS. All software-provided parameters showed low sensitivity and high specificity. The selection of new cut-off points improved the performance of all GDx parameters: VCC parameters performed better than FCC parameters; the largest AROCs were associated with the superior/nasal ratio for the GDx-FCC (0.86) and with the Number for the GDx-VCC (0.87). The LDFs provided an AROC of 0.89 with both the GDx-FCC and the GDx-VCC parameters. CONCLUSIONS. The GDx-VCC showed a higher ability in the early diagnosis of glaucoma when compared with the GDx-FCC. The individuation of the right cut-off point of selected parameters with both GDx settings performed better than the software-provided parameters, and comparably to the GDx parameters-based LDFs. CONCLUSIONS. The GDx-VCC showed a higher ability in the early diagnosis of glaucoma when compared with the GDx-FCC. The individuation of the right cut-off point of selected parameters with both GDx settings performed better than the software-provided parameters, and comparably to the GDx parameters-based LDFs.
引用
收藏
页码:468 / 476
页数:9
相关论文
共 41 条
[1]
ANDERSON D, 1999, AUTOMATED STATIC PER, V117
[2]
Nerve fiber layer measurements with scanning laser polarimetry in ocular hypertension [J].
Anton, A ;
Zangwill, L ;
Emdadi, A ;
Weinreb, RN .
ARCHIVES OF OPHTHALMOLOGY, 1997, 115 (03) :331-334
[3]
Association between scanning laser polarimetry measurements using variable corneal polarization compensation and visual in field sensitivity glaucomatous eyes [J].
Bowd, C ;
Zangwill, LM ;
Weinreb, RN .
ARCHIVES OF OPHTHALMOLOGY, 2003, 121 (07) :961-966
[4]
Bowd C, 2001, INVEST OPHTH VIS SCI, V42, P1993
[5]
BIREFRINGENCE OF THE HUMAN FOVEAL AREA ASSESSED INVIVO WITH MUELLER-MATRIX ELLIPSOMETRY [J].
BRINK, HBK ;
VANBLOKLAND, GJ .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1988, 5 (01) :49-57
[6]
Brusini P, 1996, Eur J Ophthalmol, V6, P402
[7]
Differentiating patients with glaucoma from glaucoma suspects and normal subjects by nerve fiber layer assessment with scanning laser polarimetry [J].
Choplin, NT ;
Lundy, DC ;
Dreher, AW .
OPHTHALMOLOGY, 1998, 105 (11) :2068-2076
[8]
Effect of individualized compensation for anterior segment birefringence on retinal nerve fiber layer assessments as determined by scanning laser polarimetry [J].
Choplin, NT ;
Zhou, QY ;
Knighton, RW .
OPHTHALMOLOGY, 2003, 110 (04) :719-725
[9]
Sensitivity and specificity of new GDx parameters [J].
Colen, TP ;
Tang, NEML ;
Mulder, PGH ;
Lemij, HG .
JOURNAL OF GLAUCOMA, 2004, 13 (01) :28-33
[10]
Sensitivity and specificity of the GDx: Clinical judgment of standard printouts versus the number [J].
Colen, TP ;
Lemij, HG .
JOURNAL OF GLAUCOMA, 2003, 12 (02) :129-133