Decrease in the fatigue resistance of nickel-titanium rotary instruments after clinical use in curved root canals

被引:96
作者
Bahia, MGA
Buono, VTL
机构
[1] Univ Fed Minas Gerais, Fac Dent, Dept Restorat Dent, BR-30160030 Belo Horizonte, MG, Brazil
[2] Univ Fed Minas Gerais, Sch Engn, Dept Met & Mat Engn, Belo Horizonte, MG, Brazil
来源
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY | 2005年 / 100卷 / 02期
关键词
D O I
10.1016/j.tripleo.2004.10.013
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective. The changes in fatigue resistance of nickel-titanium rotary ProFile instruments after clinical use for shaping 10 curved molar root canals were evaluated in this study. Study design. Twenty-five sets of files #20, #25, and #30 and tapers .04 and .06 were divided into 2 groups, one with 10 sets of new files that were tested in a fatigue test bench device as a control. The other, experimental group, with 15 sets of clinically used files, was tested in the same device. The Student t test was employed to compare mean values of the measured parameters. Results. A statistically significant decrease in the number of cycles to failure was determined for the clinically used files, as compared with the new ones. The fracture point was the same for all files tested. Conclusions. The clinical use of ProFile instruments for shaping curved canals reduces their fatigue resistance.
引用
收藏
页码:249 / 255
页数:7
相关论文
共 23 条
[1]  
BAHIA MGA, 2004, THESIS U FEDERAL MIN
[2]  
Courtney TH, 1990, MECH BEHAV MAT, P562
[3]   A 3-DIMENSIONAL STUDY OF CANAL CURVATURES IN THE MESIAL ROOTS OF MANDIBULAR MOLARS [J].
CUNNINGHAM, CJ ;
SENIA, ES .
JOURNAL OF ENDODONTICS, 1992, 18 (06) :294-300
[4]  
de Melo MCC, 2002, J ENDODONT, V28, P765
[5]  
DIETER GE, 1986, MECH METALLURGY, P390
[6]   Cyclic fatigue testing of ProTaper NiTi rotary instruments after clinical use [J].
Fife, D ;
Gambarini, G ;
Britto, LR .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2004, 97 (02) :251-256
[7]   Fatigue properties of a TiNi shape-memory alloy wire subjected to bending with various strain ratios [J].
Furuichi, Y ;
Tobushi, H ;
Ikawa, T ;
Matsui, R .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART L-JOURNAL OF MATERIALS-DESIGN AND APPLICATIONS, 2003, 217 (L2) :93-99
[8]   Cyclic fatigue of nickel-titanium rotary instruments after clinical use with low- and high-torque endodontic motors [J].
Gambarini, G .
JOURNAL OF ENDODONTICS, 2001, 27 (12) :772-774
[9]   Cyclic fatigue of Profile rotary instruments after prolonged clinical use [J].
Gambarini, G .
INTERNATIONAL ENDODONTIC JOURNAL, 2001, 34 (05) :386-389
[10]   Dynamic and cyclic fatigue of engine-driven rotary nickel-titanium endodontic instruments [J].
Haïkel, Y ;
Serfaty, R ;
Bateman, G ;
Senger, B ;
Allemann, C .
JOURNAL OF ENDODONTICS, 1999, 25 (06) :434-440