Increased bone contact to a calcium-incorporated oxidized commercially pure titanium implant:: an in-vivo study in rabbits

被引:31
作者
Frojd, V.
Franke-Stenport, V.
Meirelles, L.
Wennerberg, A.
机构
[1] Univ Gothenburg, Sahlgrenska Acad, Dept Biomat, S-40530 Gothenburg, Sweden
[2] Univ Gothenburg, Sahlgrenska Acad, Dept Prosthodont, S-40530 Gothenburg, Sweden
[3] Malmo Univ, Gothenburg, Sweden
关键词
calcium; oxidized implants; histomorphometry; osseointegration; surface roughness; c.p. titanium implants; surface chemistry;
D O I
10.1016/j.ijom.2008.01.020
中图分类号
R78 [口腔科学];
学科分类号
1003 [口腔医学];
摘要
The aim of this study was to evaluate the bone response to an oxidized titanium implant (Ox) and a calcium-incorporated oxidized titanium implant (Ca). A blasted titanium implant (Bl) was used as control. The implants were topographically characterized using an optical interferometer and placed: one in each distal femoral metaphysis and two in each proximal tibial metaphysis in rabbits. The rabbits were killed 12 weeks after implant insertion, and the implants and their surrounding tissues were removed en bloc for histomorphometrical evaluations. Topographical evaluation revealed three different surfaces: average height deviation (S-a, mu m) values for Ca:Ox:Bl implants were 0.3:0.6:0.9, developed surface area ratios (%) 17:44:31, number of summits per mu m(2) 208:136:118, and core fluid retention index values 1.33:1.33:1.38. The mean percentages of bone contact to the implants placed in the tibia (Ca:Ox:Bl) were 47:30:34 and to the implants placed in the femur (Ca:Ox) 32:20. The mean percentages of surrounding bone area for the implants placed in the tibia were 40:47:37 and for the implants placed in the femur 43:46. A significant increase in bone contact was found for smooth (S-a < 0.5 mu m) but more densely peaked calcium-incorporated oxidized implants when compared to slightly rougher (S-a = 0.5-1.0 mu m) oxidized or blasted implants.
引用
收藏
页码:561 / 566
页数:6
相关论文
共 25 条
[1]
Albrektsson T, 2004, INT J PROSTHODONT, V17, P536
[2]
Osteoblast adhesion on biomaterials [J].
Anselme, K .
BIOMATERIALS, 2000, 21 (07) :667-681
[3]
Surface engineering of titanium thin films with silk fibroin via layer-by-layer technique and its effects on osteoblast growth behavior [J].
Cai, Kaiyong ;
Hu, Yan ;
Jandt, Klaus D. .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2007, 82A (04) :927-935
[4]
Nanostructured ceramics for biomedical implants [J].
Catledge, SA ;
Fries, MD ;
Vohra, YK ;
Lacefield, WR ;
Lemons, JE ;
Woodard, S ;
Venugopalan, R .
JOURNAL OF NANOSCIENCE AND NANOTECHNOLOGY, 2002, 2 (3-4) :293-312
[5]
Biomimetic implant coatings [J].
Eisenbarth, E. ;
Velten, D. ;
Breme, J. .
BIOMOLECULAR ENGINEERING, 2007, 24 (01) :27-32
[6]
Characterization of titanium surfaces with calcium and phosphate and osteoblast adhesion [J].
Feng, B ;
Weng, J ;
Yang, BC ;
Qu, SX ;
Zhang, XD .
BIOMATERIALS, 2004, 25 (17) :3421-3428
[7]
GORANSSON A, 2006, THESIS GOTEBORG U
[8]
Integrin expression and function on human osteoblast-like cells [J].
Gronthos, S ;
Stewart, K ;
Graves, SE ;
Hay, S ;
Simmons, PJ .
JOURNAL OF BONE AND MINERAL RESEARCH, 1997, 12 (08) :1189-1197
[9]
Amount of hydroxyl radical on calcium-ion-implanted titanium and point of zero charge of constituent oxide of the surface-modified layer [J].
Hanawa, T ;
Kon, M ;
Doi, H ;
Ukai, H ;
Murakami, K ;
Hamanaka, H ;
Asaoka, K .
JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 1998, 9 (02) :89-92
[10]
Bone response to surface-modified titanium implants: Studies on the early tissue response to machined and electropolished implants with different oxide thicknesses [J].
Larsson, C ;
Thomsen, P ;
Aronsson, BO ;
Rodahl, M ;
Lausmaa, J ;
Kasemo, B ;
Ericson, LE .
BIOMATERIALS, 1996, 17 (06) :605-616