Summary and recommendations of a National Cancer Institute workshop on issues limiting the clinical use of Monte Carlo dose calculation algorithms for megavoltage external beam radiation therapy

被引:46
作者
Fraass, BA [1 ]
Smathers, J
Deye, J
机构
[1] Univ Michigan, Dept Radiat Oncol, Ann Arbor, MI 48109 USA
[2] Univ Calif Los Angeles, Dept Radiat Oncol, Los Angeles, CA 90024 USA
[3] NCI, Radiat Res Program, Bethesda, MD 20892 USA
关键词
Monte Carlo; clinical use; verification; recommendations;
D O I
10.1118/1.1626990
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Due to the significant interest in Monte Carlo dose calculations for external beam megavoltage radiation therapy from both the research and commercial communities, a workshop was held in October 2001 to assess the status of this computational method with regard to use for clinical treatment planning.,The Radiation Research Program of the National Cancer Institute, in conjunction with the Nuclear Data and Analysis Group at the Oak Ridge National Laboratory, gathered a group of experts in clinical radiation therapy treatment planning and Monte Carlo dose calculations, and examined issues involved in clinical implementation of Monte Carlo dose calculation methods in clinical radiotherapy. The workshop examined the current status of Monte Carlo algorithms, the rationale for using Monte Carlo, algorithmic concerns, clinical issues, and verification methodologies. Based on these discussions, the workshop developed recommendations for future NCI-funded research and development efforts. This paper briefly summarizes the issues presented at the workshop and the recommendations developed by the group. (C) 2003 American Association of Physicists in Medicine.
引用
收藏
页码:3206 / 3216
页数:11
相关论文
共 70 条
[1]   MONTE-CARLO TECHNIQUES IN MEDICAL RADIATION PHYSICS [J].
ANDREO, P .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (07) :861-920
[2]  
[Anonymous], 1989, TISS SUBST RAD DOS M
[3]   The impact of electron transport on the accuracy of computed dose [J].
Arnfield, MR ;
Siantar, CH ;
Siebers, J ;
Garmon, P ;
Cox, L ;
Mohan, R .
MEDICAL PHYSICS, 2000, 27 (06) :1266-1274
[4]  
Berger M.J., 1963, METHODS COMPUT PHYS, P135
[5]  
BERGER MJ, 1973, CCC107 ORNL
[6]   A virtual source model for Monte Carlo modeling of arbitrary intensity distributions [J].
Chetty, I ;
DeMarco, JJ ;
Solberg, TD .
MEDICAL PHYSICS, 2000, 27 (01) :166-172
[7]   Photon beam relative dose validation of the DPM Monte Carlo code in lung-equivalent media [J].
Chetty, IJ ;
Charland, PM ;
Tyagi, N ;
McShan, DL ;
Fraass, BA ;
Bielajew, AF .
MEDICAL PHYSICS, 2003, 30 (04) :563-573
[8]   Experimental validation of the DPM Monte Carlo code using minimally scattered electron beams in heterogeneous media [J].
Chetty, IJ ;
Moran, JM ;
Nurushev, TS ;
McShan, DL ;
Fraass, BA ;
Wilderman, SJ ;
Bielajew, AF .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (11) :1837-1851
[9]   CT-based Monte Carlo simulation tool for dosimetry planning and analysis [J].
DeMarco, JJ ;
Solberg, TD ;
Smathers, JB .
MEDICAL PHYSICS, 1998, 25 (01) :1-11
[10]   Efficient sampling algorithms for Monte Carlo based treatment planning [J].
DeMarco, JJ ;
Solberg, TD ;
Chetty, I ;
Smathers, JB .
RADIATION PHYSICS AND CHEMISTRY, 1998, 53 (03) :229-234