Lowering the detection limit of solvent polymeric ion-selective electrodes. 1. Modeling the influence of steady-state ion fluxes

被引:239
作者
Sokalski, T
Zwickl, T
Bakker, E
Pretsch, E
机构
[1] ETH Zurich, Dept Organ Chem, CH-8092 Zurich, Switzerland
[2] Auburn Univ, Dept Chem, Auburn, AL 36849 USA
关键词
D O I
10.1021/ac980944v
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The processes determining the lower detection limit of carrier-based ion-selective electrodes (ISEs) are described by a steady-state ion flux model under zero-current conditions. Ion-exchange and coextraction equilibria on both sides of the membrane induce concentration gradients within the organic phase and, through the resulting ion fluxes, influence the lower detection limit. The latter is shown to improve considerably when very small gradients of decreasing primary ion concentration toward the inner electrolyte solution are created. By merely altering the concentration of the inner electrolyte, detection limits may vary by more than 5 orders of magnitude. Very large gradients, however, are predicted to lead to significant depletion of analyte ions in the outer membrane surface layer and thus to apparent super-Nernstian response. The currently recommended IUPAC definition of the lower detection limit leads to nonrealistic values in such cases. Small changes in the concentration profiles within the membrane may have large effects on the response of the ISE at submicromolar levels and enhance its sensitivity to interferences during trace determinations. The model studies presented here demonstrate that trace level measurements with ISEs are feasible but often require higher membrane selectivities than expected from the Nicolskii equation.
引用
收藏
页码:1204 / 1209
页数:6
相关论文
共 24 条