Image registration of BANG® gel dose maps for quantitative dosimetry verification

被引:42
作者
Meeks, SL
Bova, FJ
Maryanski, MJ
Kendrick, LA
Ranade, MK
Buatti, JM
Friedman, WA
机构
[1] Univ Florida, Hlth Sci Ctr, Dept Radiat Oncol, Coll Med, Gainesville, FL 32610 USA
[2] Univ Florida, Coll Med, Dept Neurosurg, Gainesville, FL 32610 USA
[3] Sofamor Danek Grp, Memphis, TN USA
[4] MGS Res Inc, Guildford, Surrey, England
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1999年 / 43卷 / 05期
关键词
polymer gel; radiation dosimetry; image correlation;
D O I
10.1016/S0360-3016(98)00536-7
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: The BANG(R) (product symbol SGEL, MGS Research Inc., Guilford, CT) polymer gel has been shown to be a valuable dosimeter for determining three-dimensional (3D) dose distributions. Because the proton relaxation rate (R2) of the gel changes as a function of absorbed dose, MR scans of the irradiated gel can be used to generate 3D dose maps. Previous work with the gel, however, has not relied on precise localization of the measured dose distribution. This has limited its quantitative use, as no precise correlation exists with the planned distribution. This paper reports on a technique for providing this correlation, thus providing a quality assurance tool that includes all of the steps of imaging, treatment planning, dose calculation, and treatment localization. Methods and Materials: The BANG(R) gel formulation was prepared and poured into spherical flasks (15.3-cm inner diameter). A stereotactic head ring was attached to each flask. Three magnetic resonance imaging (MRI) and computed tomography (CT) compatible fiducial markers were placed on the flask, thus defining the central axial plane. A high-resolution CT scan was obtained of each Bask. These images were transferred to a radiosurgery treatment-planning program, where treatment plans were developed. The gels were irradiated using our systems for stereotactic radiosurgery or fractionated stereotactic radiotherapy. The gels were MR imaged, and a relative 3D dose map was created from an R2 map of these images. The dose maps were transferred to an image-correlation program, and then fused to the treatment-planning CT scan through a rigid body match of the MRI/CT-compatible fiducial markers. The fused dose maps were imported into the treatment-planning system for quantitative comparison with the calculated treatment plans. Results: Calculated and measured isodose surfaces agreed to within 2 mm at the worst points within the in-plane dose distributions. This agreement is excellent, considering that the pixel resolution of the MRI dose maps is 1.56 x 1.56 mm, and the treatment-planning dose distributions were calculated on a l-mm dose grid. All points within the dose distribution were well within the tolerances set forth for commissioning and quality assurance of stereotactic treatment-planning systems. Moreover, the quantitative evaluation presented here tests the accuracy of the entire treatment-planning and delivery process, including stereotactic frame rigidity, CT localization, CT/MR correlation, dose calculation, and radiation delivery. Conclusion: BANG(R) polymer gel dosimetry coupled with image correlation provides quantitative verification of the accuracy of 3D dose distributions. Such quantitative evaluation is imperative to ensure the high quality of the 3D dose distributions generated and delivered by stereotactic and other conformal irradiation systems. (C) 1999 Elsevier Science Inc.
引用
收藏
页码:1135 / 1141
页数:7
相关论文
共 18 条
[1]  
ATTIX F. H., 1986, Introduction to radiological physics and radiation dosimetry
[2]  
Ayyangar K M, 1998, Med Dosim, V23, P161, DOI 10.1016/S0958-3947(98)00012-0
[3]  
BARTH NH, 1990, INT J RADIAT ONCOL, V18, P425, DOI 10.1016/0360-3016(90)90111-V
[4]   The university of Florida frameless high-precision stereotactic radiotherapy system [J].
Bova, FJ ;
Buatti, JM ;
Friedman, WA ;
Mendenhall, WM ;
Yang, CC ;
Liu, C .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (04) :875-882
[5]  
BOVA FJ, IN PRESS IMAGING RAD
[6]   IMAGE CORRELATION TECHNIQUES IN RADIATION-THERAPY TREATMENT PLANNING [J].
CHEN, GTY ;
PELIZZARI, CA .
COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 1989, 13 (03) :235-240
[7]   THE UNIVERSITY-OF-FLORIDA RADIOSURGERY SYSTEM [J].
FRIEDMAN, WA ;
BOVA, FJ .
SURGICAL NEUROLOGY, 1989, 32 (05) :334-342
[8]  
FRIEDMAN WA, 1998, LINAC RADIOSURGERY P, P1
[9]   Three-dimensional visualization and measurement of conformal dose distributions using magnetic resonance imaging of BANG polymer gel dosimeters [J].
Ibbott, GS ;
Maryanski, MJ ;
Eastman, P ;
Holcomb, SD ;
Zhang, YS ;
Avison, RG ;
Sanders, M ;
Gore, JC .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (05) :1097-1103
[10]   COMPREHENSIVE QA FOR RADIATION ONCOLOGY - REPORT OF AAPM RADIATION-THERAPY COMMITTEE TASK-GROUP-40 [J].
KUTCHER, GJ ;
COIA, L ;
GILLIN, M ;
HANSON, WF ;
LEIBEL, S ;
MORTON, RJ ;
PALTA, JR ;
PURDY, JA ;
REINSTEIN, LE ;
SVENSSON, GK ;
WELLER, M ;
WINGFIELD, L .
MEDICAL PHYSICS, 1994, 21 (04) :581-618