Impedimetric immunosensor for the specific label free detection of ciprofloxacin antibiotic

被引:80
作者
Ionescu, Rodica E.
Jaffrezic-Renault, Nicole [1 ]
Bouffier, Laurent
Gondran, Chantal
Cosnier, Serge
Pinacho, Daniel G.
Marco, M. -Pilar
Sanchez-Baeza, Francisco J.
Healy, Thomas
Martelet, Claude
机构
[1] Univ Lyon 1, Sci Analyt Lab, F-69622 Villeurbanne, France
[2] Univ Grenoble 1, CNRS UMR 5250, ICMG FR 2607 CNRS, Dept Chim Mol, F-38041 Grenoble, France
[3] CSIC, CIBER Bioengn, Appl Mol Receptors Grp, ES-08034 Barcelona, Spain
[4] Natl Univ Ireland Univ Coll Cork, Tyndall Natl Inst, Cork, Ireland
[5] Ecole Cent Lyon, CEGELY, F-69134 Ecully, France
关键词
ciprofloxacin; immunosensor; polypyrrole; impedance spectroscopy;
D O I
10.1016/j.bios.2007.07.014
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Impedance spectroscopy approaches combined with the immunosensor technology have been used for the determination of trace amounts of ciprofloxacin antibiotic belonging to the fluoroquinolone family. The sensor electrode was based on the immobilization of anti-ciprofloxacin antibodies by chemical binding onto a poly(pyrrole-NHS) film clectrogenerated on a solid gold substrate. The electrode surface was modified by electropolymerization of pyrrole-NHS, antibody grafting and ciprofloxacin immunoreaction. The sensitive steps of surface modification, cyclic voltammetry (CV) and atomic force microscopy (AFM) imaging have been used for electrode surface characterization. The immunoreaction of ciprofloxacin on the grafted anti-ciprofloxacin antibody directly triggers a signal via impedance spectroscopy measurements which allows the detection of extremely low concentration of 10 pg/ml ciprofloxacin. (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:549 / 555
页数:7
相关论文
共 31 条
[1]   Ciprofloxacin pharmacokinetics in dogs following oral administration [J].
Abadia, AR ;
Aramayona, JJ ;
Munoz, MJ ;
Delfina, JMP ;
Bregante, MA .
JOURNAL OF VETERINARY MEDICINE SERIES A-PHYSIOLOGY PATHOLOGY CLINICAL MEDICINE, 1995, 42 (08) :505-511
[2]  
[Anonymous], REV MOL BIOTECHNOL, DOI DOI 10.1016/S1389-0352(01)00058-7
[3]   Biosensor surface characterisation: confirming multilayer immobilisation, determining coverage of the biospecies and establishing detection limits [J].
Berney, H ;
Roseingrave, P ;
Alderman, J ;
Lane, W ;
Collins, JK .
SENSORS AND ACTUATORS B-CHEMICAL, 1997, 44 (1-3) :341-349
[4]   Fluoroquinolones in animal health [J].
Brown, SA .
JOURNAL OF VETERINARY PHARMACOLOGY AND THERAPEUTICS, 1996, 19 (01) :1-14
[5]   ELECTRODE SYSTEMS FOR CONTINUOUS MONITORING IN CARDIOVASCULAR SURGERY [J].
CLARK, LC ;
LYONS, C .
ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1962, 102 (01) :29-&
[6]   Affinity biosensors based on electropolymerized films [J].
Cosnier, S .
ELECTROANALYSIS, 2005, 17 (19) :1701-1715
[7]   INCORPORATION OF ANIONIC COBALT PORPHYRIN BY ANION-EXCHANGE INTO POLYPYRROLE FILMS CONTAINING ALKYLAMMONIUM GROUPS [J].
DEGREGORI, I ;
CARRIER, M ;
DERONZIER, A ;
MOUTET, JC ;
BEDIOUI, F ;
DEVYNCK, J .
JOURNAL OF THE CHEMICAL SOCIETY-FARADAY TRANSACTIONS, 1992, 88 (11) :1567-1572
[8]   Development of an electrochemical immunosensor for direct detection of interferon-γ at the attomolar level [J].
Dijksma, M ;
Kamp, B ;
Hoogvliet, JC ;
van Bennekom, WP .
ANALYTICAL CHEMISTRY, 2001, 73 (05) :901-907
[9]   The qualitative and quantitative determination of quinolones of first and second generation by capillary electrophoresis [J].
Fierens, C ;
Hillaert, S ;
Van den Bossche, W .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2000, 22 (05) :763-772
[10]   The two-terminal equivalent network of a three-terminal electrochemical cell [J].
Fletcher, S .
ELECTROCHEMISTRY COMMUNICATIONS, 2001, 3 (12) :692-696