Failure mode of suture anchors as a function of insertion depth

被引:55
作者
Bynum, CK
Lee, S
Mahar, A
Tasto, J
Pedowitz, R
机构
[1] Rocky Mt Orthopaed Assoc, San Diego, CA USA
[2] Univ Calif San Diego, San Diego, CA 92103 USA
[3] Childrens Hosp, Orthoped Biomech Res Ctr, San Diego, CA USA
关键词
rotator cuff; repair; suture anchors; insertion depth; biomechanics;
D O I
10.1177/0363546504271746
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Surgeons can control not only the angle but also the depth of suture anchor placement during arthroscopic rotator cuff repair, although the tendency may be to place suture anchors on the deep side to avoid damage from prominent anchor eyelets. However, little information is available regarding possible effects of suture anchor depth on construct failure mechanisms. Hypothesis: Anchor depth affects the mode of suture failure with physiologically relevant cyclic loads. Study Design: Controlled laboratory study. Methods: Metallic screw-in suture anchors loaded with No. 2 braided polyester sutures were inserted into the bovine infraspinatus footprint with the eyelet proud, standard, or deep. Sutures were hand tied to create a closed loop. Constructs were cyclically loaded from 10 to 90 N and, if still intact at 500 cycles, taken to ultimate failure (maximum load). Results: When clinical failure was defined as greater than 3-mm construct elongation, anchors placed with the eyelet deep experienced statistically earlier clinical failure via cutting of the suture through the bone (P < .02). However, anchors placed at this level did not experience catastrophic failure during cyclic loading. The standard and proud anchors experienced 3 mm of elongation at a greater number of cycles, but the suture material degraded at the anchor eyelet, and a majority of these constructs broke during cyclic physiologic loading. At failure testing, the deep anchors had a significantly increased failure load (164 N) compared to standard (133 N) (P < .04) and proud (113 N) anchors (P < .005). Conclusion: Varying the depth of suture anchor insertion changes the mechanical properties and mode of failure of suture anchor constructs. Clinical Relevance: Surgeons should be aware of the effects of suture anchor depth and abrasive eyelet wear on construct failure during arthroscopic rotator cuff repair.
引用
收藏
页码:1030 / 1034
页数:5
相关论文
共 15 条
[1]
Suture anchors - Update 1999 [J].
Barber, FA ;
Herbert, MA .
ARTHROSCOPY, 1999, 15 (07) :719-725
[2]
SUTURE ANCHOR FAILURE STRENGTH - AN IN-VIVO STUDY [J].
BARBER, FA ;
CAWLEY, P ;
PRUDICH, JF .
ARTHROSCOPY, 1993, 9 (06) :647-652
[3]
Cyclic loading of anchor-based rotator cuff repairs: Confirmation of the tension overload phenomenon and comparison of suture anchor fixation with transosseous fixation [J].
Burkhart, SS ;
Pagan, JLD ;
Wirth, MA ;
Athanasiou, KA .
ARTHROSCOPY, 1997, 13 (06) :720-724
[4]
Cyclic loading of transosseous rotator cuff repairs: Tension overload as a possible cause of failure [J].
Burkhart, SS ;
Johnson, TC ;
Wirth, MA ;
Athanasiou, KA .
ARTHROSCOPY, 1997, 13 (02) :172-176
[5]
THE DEADMAN THEORY OF SUTURE ANCHORS - OBSERVATIONS ALONG A SOUTH TEXAS FENCE LINE [J].
BURKHART, SS .
ARTHROSCOPY, 1995, 11 (01) :119-123
[6]
Knot security in simple sliding knots and its relationship to rotator cuff repair: How secure must the knot be? [J].
Burkhart, SS ;
Wirth, HA ;
Simonich, M ;
Salem, D ;
Lanctot, D ;
Athanasiou, K .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2000, 16 (02) :202-207
[7]
Cyclic loading of rotator cuff repairs: A comparison of bioabsorbable tacks with metal suture anchors and transosseous sutures [J].
Goradia, VK ;
Mullen, DJ ;
Boucher, HR ;
Parks, BG ;
O'Donnell, JB .
ARTHROSCOPY, 2001, 17 (04) :360-364
[8]
REPAIRS OF THE ROTATOR CUFF - CORRELATION OF FUNCTIONAL RESULTS WITH INTEGRITY OF THE CUFF [J].
HARRYMAN, DT ;
MACK, LA ;
WANG, KY ;
JACKINS, SE ;
RICHARDSON, ML ;
MATSEN, FA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1991, 73A (07) :982-989
[9]
ARTHROSCOPICALLY ASSISTED ROTATOR CUFF REPAIR - CORRELATION OF FUNCTIONAL RESULTS WITH INTEGRITY OF THE CUFF [J].
LIU, SH ;
BAKER, CL .
ARTHROSCOPY, 1994, 10 (01) :54-60
[10]
Optimizing arthroscopic knots using braided or monofilament suture [J].
Loutzenheiser, TD ;
Harryman, DT ;
Ziegler, DW ;
Yung, SW .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 1998, 14 (01) :57-65