Cyclic loading of rotator cuff repairs: A comparison of bioabsorbable tacks with metal suture anchors and transosseous sutures

被引:96
作者
Goradia, VK [1 ]
Mullen, DJ [1 ]
Boucher, HR [1 ]
Parks, BG [1 ]
O'Donnell, JB [1 ]
机构
[1] Union Mem Hosp, Dept Orthopaed Surg, Baltimore, MD USA
来源
ARTHROSCOPY | 2001年 / 17卷 / 04期
关键词
rotator cuff tears; absorbable anchors; metal anchors; cyclic loading; cadaveric model; bone density;
D O I
10.1053/jars.2001.21243
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose: The purposes of the study were (1) to compare rotator cuff repair strengths after cyclic loading of 2 bioabsorbable nonsuture-based tack-type anchors, transosseous sutures, and a metal suture-based anchor, and (2) to correlate bone mineral density with mode of failure and cycles to failure. We hypothesized that specimens with a lower bone density would fail through bone at a lower number of cycles independent of the method of cuff fixation. Type of Study: Ex vivo biomechanical study. Methods: Standardized full-thickness rotator cuff defects were created in 30 fresh-frozen cadaveric shoulders that were randomized to 1 of 4 repair groups: transosseous sutures; Mitek Super suture anchors (Mitek Surgical Products, Westwood, MA); smooth bioabsorbable 8-mm Suretacs (Acufex, Smith & Nephew Endoscopy, Mansfield, MA); or spiked bioabsorbable 8-mm Suretacs (Acufex). All repairs were cyclically loaded from IO to 180 N; the numbers of cycles to 50% (gap, 5 mm) and 100% (gap, 10 mm) failure were recorded. Results: In comparing the repair groups, we found only I significant difference: the number of cycles to 100% failure was significantly higher (P < .05) for the smooth bioabsorbable tack than for the transosseous suture group. There were no statistically significant (P <less than or equal to> .05) differences in bone mineral densities with regard to each specimen's mode of failure. Conclusions: Our results suggested that immediate postoperative fixation provided by bioabsorbable tacks was similar to that provided by Mitek anchors and more stable than that provided by transosseous sutures. Therefore, the immediate postoperative biomechanical strength of bioabsorbable tacks seems comparatively adequate for fixation of selected small rotator cuff tears. However, additional evaluation in an animal model to examine degradation characteristics and sustained strength of repair is recommended before clinical use.
引用
收藏
页码:360 / 364
页数:5
相关论文
共 18 条
[1]
Orthopaedic applications for PLA-PGA biodegradable polymers [J].
Athanasiou, KA ;
Agrawal, CM ;
Barber, FA ;
Burkhart, SS .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 1998, 14 (07) :726-737
[2]
Internal fixation strength of suture anchors - Update 1997 [J].
Barber, FA ;
Herbert, MA ;
Click, JN .
ARTHROSCOPY, 1997, 13 (03) :355-362
[3]
The relationship of suture anchor failure and bone density to proximal humerus location: A cadaveric study [J].
Barber, FA ;
Feder, SM ;
Burkhart, SS ;
Ahrens, J .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 1997, 13 (03) :340-345
[4]
Suture anchor strength revisited [J].
Barber, FA ;
Herbert, MA ;
Click, JN .
ARTHROSCOPY, 1996, 12 (01) :32-38
[5]
BOSTMAN OM, 1992, CLIN ORTHOP RELAT R, P263
[6]
Cyclic loading of anchor-based rotator cuff repairs: Confirmation of the tension overload phenomenon and comparison of suture anchor fixation with transosseous fixation [J].
Burkhart, SS ;
Pagan, JLD ;
Wirth, MA ;
Athanasiou, KA .
ARTHROSCOPY, 1997, 13 (06) :720-724
[7]
Cyclic loading of transosseous rotator cuff repairs: Tension overload as a possible cause of failure [J].
Burkhart, SS ;
Johnson, TC ;
Wirth, MA ;
Athanasiou, KA .
ARTHROSCOPY, 1997, 13 (02) :172-176
[8]
Burkhart SS, 1998, AAOS INSTR COURS LEC, V47, P43
[9]
Strength of fixation with transosseous sutures in rotator cuff repair [J].
Caldwell, GL ;
Warner, JJP ;
Miller, MD ;
Boardman, D ;
Towers, J ;
Debski, R .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1997, 79A (07) :1064-1068
[10]
Craft D V, 1996, J Shoulder Elbow Surg, V5, P32, DOI 10.1016/S1058-2746(96)80028-0