Preliminary measurements of an enlarged blood pump model

被引:8
作者
Chua, LP
Akamatsu, T
Chan, WK
机构
[1] Nanyang Technol Univ, Sch Mech & Prod Engn, Singapore 639798, Singapore
[2] Setsunan Univ, Dept Mech Engn, Osaka, Japan
关键词
D O I
10.1016/S0735-1933(98)00096-7
中图分类号
O414.1 [热力学];
学科分类号
摘要
The centrifugal blood pump with a magnetically suspended impeller has shown its superiority over other artificial hearts. However, there is still insufficient understanding of fluid mechanics related issues in the clearance gap. The design requires sufficient washout in the clearance between the impeller and stationary surfaces. As the gap is only 0.2mm in width, it is very difficult to conduct measurements with present instrumentation. A 5:1 enlarged pump model was thus designed and constructed according to specifications. Flow characteristics and gap pressure measurements of the model are very close to the prototype. (C) 1998 Elsevier Science Ltd
引用
收藏
页码:1055 / 1064
页数:10
相关论文
共 7 条
[1]  
*ASME, 1989, FLUID MET THEIR THEO
[2]  
CHEREMISINOFF N, 1981, FLUID FLOW PUMPS PIP
[3]  
CHUA LP, 1996, 1 JSPS NUS SEM INT E, P175
[4]  
Gerhart P.M., 1992, FUNDAMENTALS FLUID M
[5]  
OCONNOR L, 1991, MECH ENG JUL, P37
[6]   THE ARTIFICIAL-HEART - HISTORY AND CURRENT STATUS [J].
PHILLIPS, WM .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1993, 115 (04) :555-557
[7]   Fluid dynamics of turbo pumps for artificial hearts [J].
Yamane, T ;
Ikeda, T ;
Orita, T ;
Tateishi, T .
MATERIALS SCIENCE & ENGINEERING C-BIOMIMETIC MATERIALS SENSORS AND SYSTEMS, 1996, 4 (02) :99-106