Label free gravimetric detection of epidermal growth factor receptor by antibody immobilization on quartz crystal microbalance

被引:29
作者
Chen, Jung-Chih [1 ]
Sadhasivam, S. [1 ]
Lin, Feng-Huei [1 ]
机构
[1] Natl Taiwan Univ, Inst Biomed Engn, Taipei 10764, Taiwan
关键词
Quartz crystal microbalance; Immunosensor; Epidermal growth factor receptor; Self-assembled monolayer; PATHOGENIC BACTERIA; IMMUNOSENSOR; SURFACES; BIOSENSOR; LIQUID;
D O I
10.1016/j.procbio.2010.10.006
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
An efficient and simple method for the detection of epidermal growth factor receptor (EGFR) using anti-EGFR immobilized quartz crystal microbalance (QCM) was developed. The immobilization of anti-EGFR on the gold surface of QCM electrode was constructed by a self-assembled monolayer (SAM) of thiosalicylic acid followed by the activation of 1-[3-(dimethylamino)propyl]-3-ethylcarbodiimide hydrochloride (EDC) and N-hydroxysuccinimide (NHS). In order to assess the sensitivity of QCM, two analytical procedures namely, dip-and-dry and flow cell methods were investigated. Two different antibody immobilization patterns such as chemical cross-linking and site-specific via protein G were attempted. Accordingly, protein G mediated antibody immobilization controls the quantity and orientation of the antibody molecules on the electrode surface for its high affinity to antigens. Thus, a similar immobilization strategy was utilized for the flow-cell analysis of real-time binding of EGFR. A linear relationship was observed between the frequency shift and different EGFR concentrations (0.01-10 mu g mL(-1)). The stepwise assembly of the immunosensor was characterized by quartz crystal microbalance. It was also demonstrated that, the prepared sensor surface was stable enough to withstand the repeated surface regeneration with 0.1 M NaOH. (C) 2010 Elsevier Ltd. All rights reserved.
引用
收藏
页码:543 / 550
页数:8
相关论文
共 28 条
[1]  
AKERSTROM L, 1986, J BIOL CHEM, V261, P10240
[2]  
Arteaga CL, 2002, ONCOLOGIST, V7, P31
[3]   Immobilization of antibodies through the surface regions having the highest density in lysine groups on finally inert support surfaces [J].
Batalla, Pilar ;
Mateo, Cesar ;
Grazu, Valeria ;
Fernandez-Lafuente, Roberto ;
Guisan, Jose M. .
PROCESS BIOCHEMISTRY, 2009, 44 (03) :365-368
[4]   Detection of the shrimp pathogenic bacteria, Vibrio harveyi, by a quartz crystal microbalance-specific antibody based sensor [J].
Buchatip, Sutawan ;
Ananthanawat, Cheeraporn ;
Sithigorngul, Paisarn ;
Sangvanich, Polkit ;
Rengpipat, Sirirat ;
Hoven, Voravee P. .
SENSORS AND ACTUATORS B-CHEMICAL, 2010, 145 (01) :259-264
[5]  
Curie J., 1880, C. R. Acad. Sci, V91, P294
[6]   Quartz crystal microbalance detection of glutathione-protected nanoclusters using antibody recognition [J].
Gerdon, AE ;
Wright, DW ;
Cliffel, DE .
ANALYTICAL CHEMISTRY, 2005, 77 (01) :304-310
[7]   Ni(OH)2 versus Ni/Al layered double hydroxides as matrices to immobilize glucose oxidase [J].
Guadagnini, Lorella ;
Mignani, Adriana ;
Scavetta, Erika ;
Tonelli, Domenica .
ELECTROCHIMICA ACTA, 2010, 55 (03) :1217-1220
[8]   Development of QCM biosensor to detect a marine derived pathogenic bacteria Edwardsiella tarda using a novel immobilisation method [J].
Hong, Sung-Rok ;
Choi, Suk-Jung ;
Do Jeong, Hyun ;
Hong, Suhee .
BIOSENSORS & BIOELECTRONICS, 2009, 24 (06) :1635-1640
[9]   Controlled antibody immobilization onto immunoanalytical platforms by synthetic peptide [J].
Jung, Yongwon ;
Kang, Hyo Jin ;
Lee, Jeong Min ;
Jung, Sun Ok ;
Yun, Wan Soo ;
Chung, Sang J. ;
Chung, Bong Hyun .
ANALYTICAL BIOCHEMISTRY, 2008, 374 (01) :99-105
[10]   Evaluation of a high-affinity QCM immunosensor using antibody fragmentation and 2-methacryloyloxyethyl phosphorylcholine (MPC) polymer [J].
Kurosawa, S ;
Nakamura, M ;
Park, JW ;
Aizawa, H ;
Yamada, K ;
Hirata, M .
BIOSENSORS & BIOELECTRONICS, 2004, 20 (06) :1134-1139