Visualization of dedicated catheters using fast scanning techniques with potential for MR-guided vascular interventions

被引:130
作者
Bakker, CJ
Hoogeveen, RM
Weber, J
vanVaals, JJ
Viergever, MA
Mali, WP
机构
[1] CORDIS EUROPA NV,RODEN,NETHERLANDS
[2] PHILIPS MED SYST,BEST,NETHERLANDS
关键词
MR-guided interventions; MR fluoroscopy; magnetic susceptibility; vascular interventions; catheterization;
D O I
10.1002/mrm.1910360603
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This paper reports the development of dedicated catheters and real-time MR scan techniques for guiding vascular interventional procedures. By way of phantom experiments, it is shown that proper modification of the magnetic properties of catheters allows their conspicuous and consistent visualization in subsecond 2D gradient echo images and phase contrast angiograms. Dynamic scan times as low as 0.5 s could be achieved by exploiting the keyhole technique and purposeful postprocessing. The temporal resolution and spatial resolution of the resultant scan protocol shows promise for adequate tracking of catheter manipulation.
引用
收藏
页码:816 / 820
页数:5
相关论文
共 15 条
  • [1] ACKERMAN JL, P SMRM 5 ANN M MONTR, P1131
  • [2] ANZAI Y, 1993, RADIOGRAPHICS, V113, P897
  • [3] SIMULATION OF SUSCEPTIBILITY ARTIFACTS IN 2D AND 3D FOURIER-TRANSFORM SPIN-ECHO AND GRADIENT-ECHO MAGNETIC-RESONANCE-IMAGING
    BAKKER, CJG
    BHAGWANDIEN, R
    MOERLAND, MA
    RAMOS, LMP
    [J]. MAGNETIC RESONANCE IMAGING, 1994, 12 (05) : 767 - 774
  • [4] SUSCEPTIBILITY ARTIFACTS IN 2DFT SPIN-ECHO AND GRADIENT-ECHO IMAGING - THE CYLINDER MODEL REVISITED
    BAKKER, CJG
    BHAGWANDIEN, R
    MOERLAND, MA
    FUDERER, M
    [J]. MAGNETIC RESONANCE IMAGING, 1993, 11 (04) : 539 - 548
  • [5] REAL-TIME POSITION MONITORING OF INVASIVE DEVICES USING MAGNETIC-RESONANCE
    DUMOULIN, CL
    SOUZA, SP
    DARROW, RD
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1993, 29 (03) : 411 - 415
  • [6] JOLESZ FA, 1994, MAGN RESON QUART, V10, P85
  • [7] INTRAVASCULAR MR TRACKING CATHETER - PRELIMINARY EXPERIMENTAL EVALUATION
    LEUNG, DA
    DEBATIN, JF
    WILDERMUTH, S
    MCKINNON, GC
    HOLTZ, D
    DUMOULIN, CL
    DARROW, RD
    HOFMANN, E
    VONSCHULTHESS, GK
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1995, 164 (05) : 1265 - 1270
  • [8] LUEDEKE K M, 1985, Magnetic Resonance Imaging, V3, P329
  • [9] Lufkin R B, 1990, Top Magn Reson Imaging, V2, P76
  • [10] INTERVENTIONAL MR-IMAGING
    LUFKIN, RB
    [J]. RADIOLOGY, 1995, 197 (01) : 16 - 18