Efficiency of nano-electrospray ionization

被引:122
作者
El-Faramawy, A
Siu, KWM
Thomson, BA
机构
[1] York Univ, Dept Chem, Toronto, ON M3J 2R7, Canada
[2] York Univ, Ctr Res Mass Spectrometry, Toronto, ON M3J 2R7, Canada
关键词
D O I
10.1016/j.jasms.2005.06.011
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The efficiency of nano-electrospray ionization, defined as the flux of ions reaching the detector of a triple-quadrupole mass spectrometer divided by the flux of analyte ions leaving the needle, has been measured in a series of controlled experiments with dodecyltrimethyl ammonium (DDTMA) bromide, myoglobin, Glu-[1]-fibrinopeptide, and gramicidin S. By varying the flow rate from each needle, the optimum efficiency was determined. In general, the efficiency increased as the flow rate decreased. For DDTMA, efficiencies of up to 12% were measured, although efficiencies of similar to 1% were more common. Ion current measurements indicated efficient transfer of ions from the needle through to the detector. Significant needle-to-needle variations in efficiency were encountered and attributed to variations in ion-generation efficiency.
引用
收藏
页码:1702 / 1707
页数:6
相关论文
共 12 条
[1]   THE CURRENT EMITTED BY HIGHLY CONDUCTING TAYLOR CONES [J].
DELAMORA, JF ;
LOSCERTALES, IG .
JOURNAL OF FLUID MECHANICS, 1994, 260 :155-184
[2]   Tuning of an electrospray ionization source for maximum peptide-ion transmission into a mass spectrometer [J].
Geromanos, S ;
Freckleton, G ;
Tempst, P .
ANALYTICAL CHEMISTRY, 2000, 72 (04) :777-790
[3]   On the evaporation of small ions from charged droplets [J].
IRIBARNE, JV ;
THOMSON, BA .
JOURNAL OF CHEMICAL PHYSICS, 1976, 64 (06) :2287-2294
[4]   Electrical equivalence of electrospray ionization with conducting and nonconducting needles [J].
Jackson, GS ;
Enke, CG .
ANALYTICAL CHEMISTRY, 1999, 71 (17) :3777-3784
[5]   Nanoelectrospray -: More than just a minimized-flow electrospray ionization source [J].
Juraschek, R ;
Dülcks, T ;
Karas, M .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1999, 10 (04) :300-308
[6]   Nano-electrospray ionization mass spectrometry:: addressing analytical problems beyond routine [J].
Karas, M ;
Bahr, U ;
Dülcks, T .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2000, 366 (6-7) :669-676
[7]  
Ke F, 1997, J MASS SPECTROM, V32, P992
[8]  
Kebarle P. A., 1993, ANAL CHEM, V65, P972
[9]   Effect of different solution flow rates on analyte ion signals in nano-ESI MS, or:: When does ESI turn into nano-ESI? [J].
Schmidt, A ;
Karas, M ;
Dülcks, T .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2003, 14 (05) :492-500
[10]   Declustering and fragmentation of protein ions from an electrospray ion source [J].
Thomson, BA .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1997, 8 (10) :1053-1058