Validation of a new immunoblot assay (LiaTek HIV III) for confirmation of human immunodeficiency virus infection

被引:6
作者
Zaaijer, HL
van Rixel, GACM
Kromosoeto, JNR
Balgobind-Ramdas, DR
Cuypers, HTM
Lelie, PN
机构
[1] Free Univ Amsterdam Hosp, NL-1007 MB Amsterdam, Netherlands
[2] Netherlands Red Cross, Blood Transfus Serv, Cent Lab, Dept Viral Serol, Amsterdam, Netherlands
关键词
D O I
10.1046/j.1537-2995.1998.38898375518.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND: Used as a supplemental assay, new anti-human immunodeficiency virus (HIV) immunoblots, employing recombinant and synthetic antigens, appeared to resolve the majority of samples with false-reactive Western blot results. Would it be possible to completely replace the Western blot by an immunoblot for confirmation and exclusion of HIV infection? STUDY DESIGN AND METHODS: The sensitivity of the new LiaTek HIV III immunoblot assay (Organon Teknika, Turnhout, Belgium) was tested on 416 Western-blot positive samples (386 HIV-1, 22 HIV-2, 1 HIV-1/2, and 7 HIV-O) and on 45 HIV-1 seroconversion samples. The specificity was tested on 146 samples from noninfected donors with false-positive results on a HIV screening test. RESULTS: All Western-blot-positive samples tested positive in the immunoblot (sensitivity: 100%). The immunoblot could not discriminate between HIV-1 and HIV-2 infection in 22 of 416 (5%) samples. The LiaTek assay showed reactivity in 28 of 45 seroconversion samples, whereas the Western blot reacted in 30 of 45 seroconversion samples. With false-positive donor samples, the immunoblot was indeterminate in 10 of 146 samples (specificity: 93%), and the Western blot was indeterminate in 44 of 146 samples (specificity: 70%). CONCLUSION: Like the Western blot, the immunoblot runs the risk of missing samples that are reactive by enzyme immunoassay during the early stage of HIV infection. Nevertheless, considering its superior specificity on false-positive donor samples, it appears that the immunoblot offers a cost-effective alternative to the Western blot assay for confirmation and exclusion of HIV infection.
引用
收藏
页码:776 / 781
页数:6
相关论文
共 6 条
[1]   Frequency of human immunodeficiency virus (HIV) infection among contemporary anti-HIV-1 and anti-HIV-1/2 supplemental test-indeterminate blood donors [J].
Busch, MP ;
Kleinman, SH ;
Williams, AE ;
Smith, JW ;
Ownby, HE ;
Laycock, ME ;
Lee, LLL ;
Pau, CP ;
Schreiber, GB .
TRANSFUSION, 1996, 36 (01) :37-44
[2]   Evaluation of Chiron HIV-1/HIV-2 recombinant immunoblot assay [J].
Kline, RL ;
McNairn, D ;
Holodniy, M ;
Mole, L ;
Margolis, D ;
Blattner, W ;
Quinn, TC .
JOURNAL OF CLINICAL MICROBIOLOGY, 1996, 34 (11) :2650-2653
[3]   False-positive human immunodeficiency virus type 1 Western blot tests in noninfected blood donors [J].
Sayre, KR ;
Dodd, RY ;
Tegtmeier, G ;
Layug, L ;
Alexander, SS ;
Busch, MP .
TRANSFUSION, 1996, 36 (01) :45-52
[4]   Use of human immunodeficiency virus (HIV) type 1 and 2 recombinant strip immunoblot assay to resolve enzyme immunoassay anti-HIV-2-repeatably reactive samples after anti-HIV-1/2 combination enzyme immunoassay screening [J].
Tobler, LH ;
Kaufman, E ;
Gefter, N ;
Schable, C ;
Busch, MP .
TRANSFUSION, 1997, 37 (09) :921-925
[5]   New anti-human immunodeficiency virus immunoblot assays resolve nonspecific Western blot results [J].
Zaaijer, HL ;
vanRixel, T ;
vanExelOehlers, P ;
Cuypers, HTM ;
Lelie, PN .
TRANSFUSION, 1997, 37 (02) :193-198
[6]   EARLY DETECTION OF ANTIBODIES TO HIV-1 BY 3RD-GENERATION ASSAYS [J].
ZAAIJER, HL ;
VONEXELOEHLERS, P ;
KRAAIJEVELD, T ;
ALTENA, E ;
LELIE, PN .
LANCET, 1992, 340 (8822) :770-772