A comparison of bone remodelling around hydroxyapatite-coated, porous-coated and grit-blasted hip replacements retrieved at post-mortem

被引:104
作者
Coathup, MJ
Blunn, GW
Flynn, N
Williams, C
Thomas, NP
机构
[1] UCL, Ctr Biomed Engn, Royal Natl Orthopaed Hosp, Stanmore HA7 4LP, Middx, England
[2] Basingstoke Dist Gen Hosp, Basingstoke RG24 9NA, Hants, England
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 2001年 / 83B卷 / 01期
关键词
D O I
10.1302/0301-620X.83B1.10062
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We investigated the implant-bone interface around one design of femoral stem, proximally coated with either a plasma-sprayed porous coating (plain porous) or a hydroxyapatite porous coating (porous HA), or which had been grit-blasted (Interlok). Of 165 patients implanted with a Bimetric hip hemiarthroplasty (Biomet, Bridgend, UK) specimens were retrieved from 58 at post-mortem, We estimated ingrowth and attachment of bone to the surface of the implant in 21 of these, eight plain porous, seven porous HA and six Interlok, using image analysis and light morphometric techniques. The amount of HA coating was also quantified. There was significantly more ingrowth (p = 0.012) and attachment of bone (p < 0.05) to the porous HA surface (mean bone ingrowth 29.093 +/- 2.019%; mean bone attachment 37.287 +/- 2.489%) than to the plain porous surface (mean bone ingrowth 21.762 +/- 2.068%; mean bone attachment 18.9411 +/- 1.971%). There was no significant difference in attachment between the plain porous and Interlok surfaces. Bone grew more evenly over the surface of the HA coating whereas on the porous surface, bone ingrowth and attachment occurred more on the distal and medial parts of the coated surface. No significant differences in the volume of HA were found with the passage of time. This study shows that HA coating increases the amount of ingrowth and attachment of bone and leads to a more even distribution of bone over the surface of the implant. This may have implications in reducing stress shielding and limiting osteolysis induced by wear particles.
引用
收藏
页码:118 / 123
页数:6
相关论文
共 22 条
[1]   HYDROXYAPATITE-COATED FEMORAL STEMS - HISTOLOGICAL ANALYSIS OF COMPONENTS RETRIEVED AT AUTOPSY [J].
BAUER, TW ;
GEESINK, RCT ;
ZIMMERMAN, R ;
MCMAHON, JT .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1991, 73A (10) :1439-1452
[2]  
BLOEBAUM RD, 1991, CLIN ORTHOP RELAT R, P97
[3]  
BOBYN JD, 1984, ORTHOPEDICS, V7, P1410, DOI 10.3928/0147-7447-19840901-05
[4]   EVIDENCE OF BONE INGROWTH INTO A POROUS-COATED PROSTHESIS - A CASE-REPORT [J].
BROOKER, AF ;
COLLIER, JP .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1984, 66A (04) :619-621
[5]   THE UNCEMENTED POROUS-COATED ANATOMIC TOTAL HIP-PROSTHESIS - 2-YEAR RESULTS OF A PROSPECTIVE CONSECUTIVE SERIES [J].
CALLAGHAN, JJ ;
DYSART, SH ;
SAVORY, CG .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (03) :337-346
[6]  
Cook S D, 1988, J Arthroplasty, V3, P249
[7]  
COOK SD, 1988, CLIN ORTHOP RELAT R, P90
[8]   EARLY CLINICAL-EXPERIENCE WITH HYDROXYAPATITE-COATED FEMORAL IMPLANTS [J].
DANTONIO, JA ;
CAPELLO, WN ;
CROTHERS, OD ;
JAFFE, WL ;
MANLEY, MT .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1992, 74A (07) :995-1008
[9]  
DONATH K, 1990, OSSEOINTEGRATED IMPL, P99
[10]   EFFECT OF HYDROXYAPATITE IMPREGNATION ON SKELETAL BONDING OF POROUS COATED IMPLANTS [J].
DUCHEYNE, P ;
HENCH, LL ;
KAGAN, A ;
MARTENS, M ;
BURSENS, A ;
MULIER, JC .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1980, 14 (03) :225-237