A GSO depth of interaction detector for PET

被引:90
作者
Yamamoto, S
Ishibashi, H
机构
[1] Kobe City Coll Technol, Nishi Ku, Kobe, Hyogo, Japan
[2] Hitachi Chem Co Ltd, Tsukuba Res Lab, Tsukuba, Ibaraki, Japan
关键词
D O I
10.1109/23.681982
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We have developed a GSO depth of interaction detector. Three GSO scintillators with different concentration of Ce were chosen and, slacked along depth of interaction direction and optically coupled to a photomultiplier tube (PMT) or a position sensitive photomultiplier tube (PSPMT). In the beginning, pulse shape, light output and zero-cross time were measured to select a proper combination of GSO scintillators with different concentration of Ce. Selecting three GSO scintillators with different concentration of Ce, a single slacked GSO/PMT depth of interaction detector was fabricated for estimating the possibility of this method. Using the zero-cross time analysis, three peaks correspond to decay limes of the three GSO scintillators were obtained. Finally, a GSO/PSPMT depth of interaction block detector was fabricated and tested. The block detector consists of 7x7 matrix of GSO scintillators which have three layers of different concentration of Ce optically coupled to a PSPMT. The pulse shape analysis was used for determination of depth of interaction direction while the Anger logic was used for transaxial and axial directions. Good separation was observed in depth of interaction direction as well as transaxial and axial directions. These results indicate that the GSO block detector may become one solution to develop a positron emission tomograph (PET) with depth of interaction detection capability.
引用
收藏
页码:1078 / 1082
页数:5
相关论文
共 10 条
[1]   PERFORMANCE OF NON-DELAY-LINE CONSTANT-FRACTION DISCRIMINATOR TIMING-CIRCUITS [J].
BINKLEY, DM .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (04) :1169-1175
[2]   Investigation of LSO crystals for high spatial resolution positron emission tomography [J].
Casey, ME ;
Eriksson, L ;
Schmand, M ;
Andreaco, MS ;
Paulus, M ;
Dahlbom, M ;
Nutt, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) :1109-1113
[3]   COLLECTION OF SCINTILLATION LIGHT FROM SMALL BGO CRYSTALS [J].
CHERRY, SR ;
SHAO, YP ;
TORNAI, MP ;
SIEGEL, S ;
RICCI, AR ;
PHELPS, ME .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (04) :1058-1063
[4]   Performance of a YSO/LSO phoswich detector for use in a PET/SPECT system [J].
Dahlbom, M ;
MacDonald, LR ;
Eriksson, L ;
Paulus, M ;
Andreaco, M ;
Casey, ME ;
Moyers, C .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) :1114-1119
[5]  
IISHIBASHI H, 1989, IEEE T NUCL SCI, V36, P170
[6]   Performance studies of a depth encoding multicrystal detector for PET [J].
Moisan, C ;
Tsang, G ;
Rogers, JG ;
Hoskinson, EM .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (03) :1926-1931
[7]   DESIGN STUDIES FOR A PET DETECTOR MODULE USING A PIN PHOTODIODE TO MEASURE DEPTH OF INTERACTION [J].
MOSES, WW ;
DERENZO, SE .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (04) :1441-1445
[8]   DESIGNING A STRATIFIED DETECTION SYSTEM FOR PET CAMERAS [J].
WONG, WH .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1986, 33 (01) :591-596
[9]   Development of a high resolution beta camera for a direct measurement of positron distribution on brain surface [J].
Yamamoto, S ;
Seki, C ;
Kashikura, K ;
Fujita, H ;
Matsuda, T ;
Ban, R ;
Kanno, I .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (04) :1538-1542
[10]  
YAMAMOTO S, 1996, IEEE NSS MIC RECORDS, V2, P1216